斜视手术前后超声生物显微镜下睑袋/球囊技术定位水平眼外肌附着点的准确性

Accuracy of the Ultrasound Biomicroscopy Bag/Balloon Technique in Locating Horizontal Extraocular Muscle Insertions Before and After Strabismus Surgery.

作者信息

Mezad-Koursh Daphna, Rosenblatt Amir, Cohen Shai, Neudorfer Meira, Keynan Idit, Stolovitch Chaim, Zur Dinah

出版信息

J Pediatr Ophthalmol Strabismus. 2020 Jan 1;57(1):12-20. doi: 10.3928/01913913-20191115-01.

Abstract

PURPOSE

To investigate the accuracy of high frequency ultrasound biomicroscopy using the bag/balloon technique to locate recti muscle insertions before and after strabismus surgery.

METHODS

This was a prospective masked study. The distance from the limbus to horizontal recti muscle insertions was measured by caliper intraoperatively and ultrasound biomicroscopy preoperatively and during follow-up. Accuracy was defined as the difference between ultrasound biomicroscopy and caliper measurements.

RESULTS

Thirty-nine muscles (19 medial rectus and 20 lateral rectus) of 22 patients were included. The mean age of the patients was 34.7 ± 15.5 years (range: 18 to 78 years). Follow-up measurements were available for 25 muscles: 12 muscles were recessed and 13 were resected/advanced. Twelve of the included muscles underwent prior surgery. There was a strong correlation between preoperative mean ultrasound biomicroscopy and caliper measurements (P < .001, r = 0.872) without significant difference between the accuracy of identification of medial rectus and lateral rectus muscles (P = .116 and .377, respectively). On average, postoperative ultrasound biomicroscopy measurements of recessed muscles were 1.15 mm greater than caliper measurements at the end of surgery (P = .003), whereas the mean difference in the resected muscles was not statistically significant (-0.07 mm, P > .999). Overall, in 16 of 25 muscles (64%), the differences between the postoperative measurements and the end of surgery measurements were less than 1 mm. There was no correlation between the accuracy and the time of ultrasound biomicroscopy after surgery (P = .516).

CONCLUSIONS

The ultrasound biomicroscopy bag/balloon technique is an accurate and reliable method of locating recti muscle insertions before and after strabismus surgery. Ultrasound biomicroscopy may assist strabismologists in planning repeated operations. [J Pediatr Ophthalmol Strabismus. 2020;57(1):12-20.].

摘要

目的

探讨使用袋/球囊技术的高频超声生物显微镜检查在斜视手术前后定位直肌附着点的准确性。

方法

这是一项前瞻性盲法研究。术中用卡尺测量角膜缘至水平直肌附着点的距离,术前及随访期间用超声生物显微镜测量。准确性定义为超声生物显微镜测量值与卡尺测量值之间的差异。

结果

纳入22例患者的39条肌肉(19条内直肌和20条外直肌)。患者的平均年龄为34.7±15.5岁(范围:18至78岁)。对25条肌肉进行了随访测量:12条肌肉进行了后徙,13条肌肉进行了切除/前徙。纳入的肌肉中有12条曾接受过手术。术前超声生物显微镜平均测量值与卡尺测量值之间存在强相关性(P <.001,r = 0.872),内直肌和外直肌识别准确性之间无显著差异(分别为P =.116和.377)。平均而言,后徙肌肉术后超声生物显微镜测量值比手术结束时卡尺测量值大1.15 mm(P =.003),而切除肌肉的平均差异无统计学意义(-0.07 mm,P>.999)。总体而言,25条肌肉中有16条(64%)术后测量值与手术结束时测量值之间的差异小于1 mm。准确性与术后超声生物显微镜检查时间之间无相关性(P =.516)。

结论

超声生物显微镜袋/球囊技术是斜视手术前后定位直肌附着点的准确可靠方法。超声生物显微镜检查可协助斜视专家规划再次手术。[《小儿眼科与斜视杂志》。2020;57(1):12 - 20。]