Mitral regurgitation volume (MRvol) by quantitative pulsed Doppler (QPD) method previously recommended suffers from geometric assumption error because of circular geometric assumption of mitral annulus (MA). Therefore, the aim of this study was to evaluate the impact of different geometric assumption of MA on the assessment of MRvol by two-dimensional transthoracic echocardiographic QPD method.
This study included 88 patients with varying degrees of mitral regurgitation (MR). The MRvol was evaluated by QPD method using circular or ellipse geometric assumption of MA. MRvol derived from effective regurgitant orifice area by real time three-dimensional echocardiography (RT3DE) multiplied by MR velocity-time integral was used as reference method.
Assumption of a circular geometry of MA, QPD-MA and QPD-MA overestimated the MRvol by a mean difference of 10.4 ml (P < 0.0001) and 22.5 ml (P < 0.0001) compared with RT3DE. Assumption of an ellipse geometry of MA, there was no significant difference of MRvol (mean difference = 1.7 ml, P = 0.0844) between the QPD-MA and the RT3DE.
Assuming that the MA was circular geometry previously recommended, the MRvol by QPD-MA was overestimated compared with the reference method. However, assuming that the MA was ellipse geometry, the MRvol by the QPD-MA has no significant difference with the reference method.
先前推荐的定量脉冲多普勒(QPD)法测量二尖瓣反流容积(MRvol)时,由于二尖瓣环(MA)采用圆形几何假设,存在几何假设误差。因此,本研究旨在评估MA不同几何假设对二维经胸超声心动图QPD法评估MRvol的影响。
本研究纳入88例不同程度二尖瓣反流(MR)患者。采用MA的圆形或椭圆形几何假设,通过QPD法评估MRvol。以实时三维超声心动图(RT3DE)测量的有效反流口面积乘以MR速度时间积分得出的MRvol作为参考方法。
与RT3DE相比,假设MA为圆形几何形状时,QPD-MA和QPD-MA高估MRvol的平均差值分别为10.4ml(P<0.0001)和22.5ml(P<0.0001)。假设MA为椭圆形几何形状时,QPD-MA与RT3DE之间的MRvol无显著差异(平均差值=1.7ml,P=0.0844)。
与参考方法相比,先前推荐假设MA为圆形几何形状时,QPD-MA高估了MRvol。然而,假设MA为椭圆形几何形状时,QPD-MA得出的MRvol与参考方法无显著差异。