The aim of this study is to perform a systematic review of the feasibility and clinical application of double low-dose CT pulmonary angiography (CTPA) in the diagnosis of patients with suspected pulmonary embolism.
A total of 13 studies were found to meet selection criteria reporting both low radiation dose (70 or 80 kVp versus 100 or 120 kVp) and low contrast medium dose CTPA protocols. Lowdose CTPA resulted in radiation dose reduction from 29.6% to 87.5% in 12 studies (range: 0.4 to 23.5 mSv), while in one study, radiation dose was increased in the dual-energy CT group when compared to the standard 120 kVp group. CTPA with use of low contrast medium volume (range: 20 to 75 ml) was compared to standard CTPA (range: 50 to 101 ml) in 12 studies with reduction between 25 and 67%, while in the remaining study, low iodine concentration was used with 23% dose reduction achieved. Quantitative assessment of image quality (in terms of signal-to-noise ratio and contrast-to-noise ratio) showed that low-dose CTPA was associated with higher, lower and no change in image quality in 3, 3 and 6 studies, respectively when compared to the standard CTPA protocol. The subjective assessment indicated similar image quality in 11 studies between low-dose and standard CTPA groups, and improved image quality in 1 study with low-dose CTPA.
This review shows that double low-dose CTPA is feasible in the diagnosis of pulmonary embolism with significant reductions in both radiation and contrast medium doses, without compromising diagnostic image quality.
本研究旨在对双低剂量CT肺动脉造影(CTPA)在疑似肺栓塞患者诊断中的可行性及临床应用进行系统评价。
共发现13项研究符合纳入标准,这些研究报告了低辐射剂量(70或80 kVp对比100或120 kVp)和低对比剂剂量的CTPA方案。12项研究中,低剂量CTPA使辐射剂量降低了29.6%至87.5%(范围:0.4至23.5 mSv),而在一项研究中,与标准120 kVp组相比,双能CT组的辐射剂量增加。在12项研究中,将使用低对比剂体积(范围:20至75 ml)的CTPA与标准CTPA(范围:50至101 ml)进行比较,对比剂减少了25%至67%,而在其余研究中,使用低碘浓度实现了23%的剂量减少。图像质量的定量评估(根据信噪比和对比噪声比)显示,与标准CTPA方案相比,低剂量CTPA在3项、3项和6项研究中分别与图像质量更高、更低和无变化相关。主观评估表明,11项研究中低剂量和标准CTPA组的图像质量相似,1项研究中低剂量CTPA的图像质量有所改善。
本综述表明,双低剂量CTPA在肺栓塞诊断中是可行的,可显著降低辐射和对比剂剂量,且不影响诊断图像质量。