Ren Zhengyi, Guo Meiru, Cheng Yongjun, Sun Wenjun, Li Gang, Dong Meng, Li Yali, Xi Zhenhua, Pei Xiaoqiang, Niu Hongzhi, Tan Guobin, Zhou Zhen, Huang Zhengxu
Science and Technology on Vacuum Technology and Physics Laboratory , Lanzhou Institute of Physics , Lanzhou 730000 , China.
Institute of Mass Spectrometer and Atmospheric Environment , Jinan University , Guangzhou 510632 , China.
J Am Soc Mass Spectrom. 2020 Feb 5;31(2):434-440. doi: 10.1021/jasms.9b00092. Epub 2020 Jan 2.
A miniature reflectron time-of-flight mass spectrometer (TOF MS) with orthogonal extraction coupled with electron impact (EI) ionization source can be used to perform in situ gas composition analysis in a planetary environment. However, performances such as the mass resolution, sensitivity, limit of detection, mass range, and mass accuracy are often decreased because of miniaturization. Herein, a compact instrument for space applications has been developed, and its performance has been evaluated. The mass of the TOF MS is 13.4 kg, with dimensions of 300 mm × 200 mm × 200 mm, and the power consumption is 25 W. In this paper, the design of the ion source, mass analyzer, and detector is discussed in detail. The upper limit of the mass range is greater than 500 amu, and the best resolving power obtained so far on the miniature TOF MS is around 405 at full width half maximum (FWHM); other performance indexes of the instrument are also determined, where the worst case for mass stability is 0.49%, together with a mass accuracy of 0.12% and a sensitivity of 0.6 mV/ppm.
一种带有正交提取装置并与电子轰击(EI)电离源相结合的微型反射式飞行时间质谱仪(TOF MS)可用于在行星环境中进行原位气体成分分析。然而,由于小型化,诸如质量分辨率、灵敏度、检测限、质量范围和质量精度等性能往往会下降。在此,已开发出一种用于空间应用的紧凑型仪器,并对其性能进行了评估。该TOF MS的质量为13.4千克,尺寸为300毫米×200毫米×200毫米,功耗为25瓦。本文详细讨论了离子源、质量分析器和检测器的设计。质量范围的上限大于500原子质量单位(amu),到目前为止,微型TOF MS在半高宽(FWHM)处获得的最佳分辨率约为405;还确定了该仪器的其他性能指标,其中质量稳定性的最坏情况为0.49%,质量精度为0.12%,灵敏度为0.6毫伏/百万分率(ppm)。