Eur Radiol. 2020 Jun;30(6):3356-3362. doi: 10.1007/s00330-020-06695-y. Epub 2020 Feb 14.
Automated ultrasound of the breast has the advantage to have the whole breast scanned by technicians. Consequently, feedback to the radiologist about concurrent focal abnormalities (e.g., palpable lesions) is lost. To enable marking of patient- or physician-reported focal abnormalities, we aimed to develop skin markers that can be used without disturbing the interpretability of the image.
Disk-shaped markers were casted out of silicone. In this IRB-approved prospective study, 16 patients were included with a mean age of 57 (39-85). In all patients, the same volume was imaged twice using an automated breast ultrasound system, once with and once without a marker in place. Nine radiologists from two medical centers filled scoring forms regarding image quality, image interpretation, and confidence in providing a diagnosis based on the images.
Marker adhesion was sufficient for automated scanning. Observer scores showed a significant shift in scores from excellent to good regarding diagnostic yield/image quality (χ, 15.99, p < 0.01), and image noise (χ, 21.20, p < 0.01) due to marker presence. In 93% of cases, the median score of observers "agree" with the statement that marker-induced noise did not influence image interpretability. Marker presence did not interfere with confidence in diagnosis (χ, 6.00, p = 0.20).
Inexpensive, easy producible skin markers can be used for accurate lesion marking in automated ultrasound examinations of the breast while image interpretability is preserved. Any marker-induced noise and decreased image quality did not affect confidence in providing a diagnosis.
• The use of a skin marker enables the reporting radiologist to identify a location which a patient is concerned about. • The developed skin marker can be used for accurate breast lesion marking in ultrasound examinations.
自动化乳腺超声具有由技术员扫描整个乳房的优势。因此,关于并发局灶性异常(例如可触及的病变)的反馈会丢失给放射科医生。为了能够标记患者或医生报告的局灶性异常,我们旨在开发可使用而不会干扰图像可解释性的皮肤标记物。
圆盘形标记物由硅酮铸造而成。在这项经过机构审查委员会批准的前瞻性研究中,纳入了 16 名平均年龄为 57 岁(39-85 岁)的患者。在所有患者中,使用自动乳腺超声系统两次对同一体积进行成像,一次有标记物,一次无标记物。来自两个医疗中心的 9 名放射科医生根据图像填写了关于图像质量、图像解释和提供诊断信心的评分表。
标记物的粘附性足以进行自动扫描。观察者评分显示,由于标记物的存在,在诊断率/图像质量(χ²,15.99,p<0.01)和图像噪声(χ²,21.20,p<0.01)方面,评分从优秀变为良好。在 93%的情况下,观察者的中位数评分“同意”标记物引起的噪声不会影响图像可解释性的说法。标记物的存在并不干扰诊断信心(χ²,6.00,p=0.20)。
廉价、易于制作的皮肤标记物可用于在自动乳腺超声检查中准确标记病变,同时保留图像可解释性。任何标记物引起的噪声和图像质量下降都不会影响提供诊断的信心。
使用皮肤标记物可以使报告放射科医生识别出患者关注的位置。
开发的皮肤标记物可用于超声检查中准确标记乳腺病变。