Department of Paediatric Cardiology, The Children's Hospital and Institute of Child Health, Lahore, Pakistan.
Division of Pediatric Cardiology, Congenital and Structural Heart Disease, Children's Hospital of Georgia, Augusta University, Augusta, GA, USA.
Cardiol Young. 2020 Mar;30(3):444-445. doi: 10.1017/S104795112000027X. Epub 2020 Feb 18.
We report a case of Figulla-II Occlutech septal occluder malposition with residual shunt at posteriosuperior margin of an atrial septal defect. Improvising its bioptome type delivery cable, same system was used to recapture the device and redeploy it successfully. This report highlights a potential malfunction of Figulla-II Occlutech disc and the advantage of its delivery system for retrieval of the device.
我们报告了一例 Figulla-II Occlutech 隔瓣型封堵器位置不良,伴有房间隔缺损后上缘残余分流。我们巧妙地利用其活检型输送导丝,成功地重新捕获并重新部署该器械。本报告强调了 Figulla-II Occlutech 瓣的潜在故障以及该器械输送系统在器械回收方面的优势。