Abstract

OBJECTIVE

Sleep apnea research cites that an oral appliance, which places the mandible in a more forward position and the genioglossus (tongue muscle) on the floor of the mouth, improves aspects of the pharyngeal opening. Exercise science research has cited improvements in airway dynamics and physiological variables with oral appliance use during exercise. Thus, the purpose of this study was to design an oral appliance that would act on the genioglossus and determine if there were effects on respiratory parameters while exercising.

MATERIALS AND METHODS

Seventeen healthy subjects ages 18-43 participated in this study. Prior to the exercise protocol, the order of the oral applicance (OA) or no oral appliance (no OA) condition was randomly assigned to subjects, with subjects completing both conditions. Respiratory parameters (respiratory rate, ventilation, oxygen, and carbon dioxide) were measured between conditions while subjects ran for 10 min at steady state.

RESULTS

The results demonstrated that both respiratory rate (25.97 BPM, OA and 28.35 BPM, no OA) and ventilation (47.66 l/min, OA and 50.34 l/min, No OA) were significantly lowered (p < .01) in the OA condition. There were no differences in carbon dioxide (1.89 l/min, no OA and 1.88 l/min, OA) or oxygen outcomes (2.17 l/min, no OA and 2.17 l/min OA).

DISCUSSION

The outcomes from this study suggest that the design of the oral appliance elicits an effect on the genioglossus, thereby resulting in lowered respiratory rate and ventilation with no negative effects on oxygen uptake during exercise.

摘要

目的

睡眠呼吸暂停研究表明,一种口腔器械可将下颌骨置于更向前的位置,并使颏舌肌位于口腔底部,从而改善咽腔开口的各个方面。运动科学研究表明,在运动过程中使用口腔器械可以改善气道动力学和生理变量。因此,本研究的目的是设计一种作用于颏舌肌的口腔器械,并确定其在运动时对呼吸参数的影响。

材料与方法

本研究纳入了 17 名年龄在 18-43 岁的健康受试者。在进行运动方案之前,根据随机分配的顺序,受试者分别接受口腔器械(OA)或无口腔器械(无 OA)条件,然后完成两种条件下的测试。在受试者以稳定状态跑步 10 分钟期间,分别测量两种条件下的呼吸参数(呼吸频率、通气量、氧气和二氧化碳)。

结果

结果表明,OA 条件下呼吸频率(25.97 次/分,OA 和 28.35 次/分,无 OA)和通气量(47.66 l/min,OA 和 50.34 l/min,无 OA)均显著降低(p <.01)。二氧化碳(1.89 l/min,无 OA 和 1.88 l/min,OA)和氧气摄取(2.17 l/min,无 OA 和 2.17 l/min,OA)结果无差异。

讨论

本研究结果表明,口腔器械的设计对颏舌肌产生了影响,从而导致呼吸频率和通气量降低,而在运动过程中对氧气摄取没有负面影响。