Lee P L, Selzer R H
Jet Propulsion Laboratory, California Institute of Technology, Pasadena 91109.
Med Phys. 1988 Jul-Aug;15(4):576-80. doi: 10.1118/1.596296.
We have reexamined the problem of estimating the luminal area of an elliptically shaped coronary artery cross section from two or more radiographic diameter measurements. The expected error is found to be much smaller than the maximum potential error. In the cases of two orthogonal views, closed form expressions have been derived for calculating the area and the uncertainty. Assuming that the underlying ellipse has limited ellipticity (major/minor axis ratio less than five), it is shown that the average uncertainty in the area is less than 14%. When more than two views are available, we suggest using a least-squares fit method to extract all available information from the data.
我们重新审视了根据两个或更多的射线照相直径测量值来估算椭圆形冠状动脉横截面管腔面积的问题。结果发现,预期误差远小于最大潜在误差。对于两个正交视图的情况,已经推导出了用于计算面积和不确定性的封闭形式表达式。假设基础椭圆的椭圆率有限(长轴/短轴比率小于5),结果表明面积的平均不确定性小于14%。当有两个以上的视图可用时,我们建议使用最小二乘法拟合方法从数据中提取所有可用信息。