Regenass S, Pfisterer M, Bassignana V, Müller-Brand J, Burkart F
Kardiologische und nuklearmedizinische Abteilungen, Kantonsspital Basel.
Schweiz Med Wochenschr. 1988 Nov 19;118(46):1719-23.
To validate a new portable scintillation probe (VEST) for determination of LV function, 68 patients were studied after standard radionuclide angiography (RNA). Reproducibility of VEST LV-ejection fraction was good (rest r = 0.93; exercise r = 0.96) with unchanged probe position, somewhat less good after probe repositioning (r = 0.84) and correlated with RNA (rest r = 0.79; exercise r = 0.83). Long term measurements in 7 patients showed valid data for up to 12 hours. Combined with simultaneous two lead ECG recording VEST seems suitable for detection of transient changes of LV function such as occur during ischemia or due to drug treatment.
为验证一种用于测定左心室功能的新型便携式闪烁探头(VEST),在标准放射性核素血管造影(RNA)后对68例患者进行了研究。在探头位置不变的情况下,VEST左心室射血分数的可重复性良好(静息状态下r = 0.93;运动状态下r = 0.96),探头重新定位后可重复性稍差(r = 0.84),且与RNA相关(静息状态下r = 0.79;运动状态下r = 0.83)。对7例患者的长期测量显示,长达12小时的数据有效。结合同步双导联心电图记录,VEST似乎适用于检测左心室功能的短暂变化,如缺血期间或药物治疗引起的变化。