Bartels P C, Roijers A F, Soons J B
Department of Clinical Chemistry and Haematology, Medical Centre Alkmaar, The Netherlands.
Scand J Clin Lab Invest. 1988 Oct;48(6):589-93. doi: 10.3109/00365518809085777.
Characteristics with respect to light scattering intensity (LSI) were measured on latex spheres and on sphered and unsphered red blood cells (RBC). From the discrepancies in LSI signals from polymer spheres and RBCs it is obvious that latices cannot be used for absolute calibration of RBC sizes. Sphering of RBCs did not give rise to a closer resemblance. Monodisperse latex spheres produce extremely narrow size distribution histograms. Regarding these sharp peaks and the long-term stability, polymer spheres offer definite advantages if compared with blood controls to which preservatives were added. A shift from the RBC size distribution histogram, caused by gradually developing instrument drift, can be easily detected and corrected at an early stage.
对乳胶球以及球形和非球形红细胞(RBC)进行了光散射强度(LSI)相关特性的测量。从聚合物球和红细胞的LSI信号差异可以明显看出,乳胶不能用于红细胞大小的绝对校准。红细胞的球形化并没有使其更相似。单分散乳胶球产生的尺寸分布直方图极其狭窄。鉴于这些尖锐的峰值和长期稳定性,与添加了防腐剂的血液对照相比,聚合物球具有明显优势。由仪器逐渐漂移导致的红细胞大小分布直方图的偏移可以很容易地在早期检测到并得到纠正。