Extracorporeal CO2-removal with a heparin coated artificial lung.

作者信息

Peters J, Radermacher P, Kuntz M E, Rosenbauer K A, Breulmann M, Bürrig K F, Hopf H B, Rossaint R, Schulte H D, Olsson P

机构信息

Department of Anesthesiology, University of Düsseldorf, FRG.

出版信息

Intensive Care Med. 1988;14(5):578-84. doi: 10.1007/BF00263533.

Abstract

Treatment of severe acute respiratory failure with extracorporeal gas exchange necessitating near complete systemic anticoagulation requires a delicate balance to be maintained between disseminated intravascular coagulation and hemorrhagic complications. The present study describes our first experience using a heparin coated extracorporeal artificial lung and circuitry during clinical extracorporeal CO2 removal. In spite of a partial thromboplastin time and activated clotting time within or close to the normal range, neither laboratory evidence for disseminated intravascular coagulation induced by the extracorporeal circuit nor thrombi in the pulmonary vasculature were found. Scanning electron microscopy of the heparin coated hollow fiber gas exchanger demonstrated only minor deposits on the surface. Use of a heparin coated artificial lung may enhance the margin of safety of extracorporeal gas exchange and ultimately broaden its indications.