Thompson D J, Robinette L N
Dorn Veterans Hospital, Columbia, South Carolina.
Ear Hear. 1988 Oct;9(5):290-5. doi: 10.1097/00003446-198810000-00010.
Temporal response to a simulated rapid change in acoustic immittance was recorded from 13 aural acoustic-immittance devices. Ten measures of latency were obtained from the responses. The derived temporal data showed considerable variance across instruments. The variance was affected by the site from which the temporal response was recorded in the instrument and in one case by differences among samples of the same model.