OCT-A 相关的视网膜前膜的分段错误和运动伪影。

Segmentation errors and motion artifacts in OCT-A associated with epiretinal membranes.

机构信息

Department of Ophthalmology, "Korgialenio-Benakio" General Hospital, Athens, Greece.

Department of Ophthalmology, "Korgialenio-Benakio" General Hospital, Athens, Greece.

出版信息

Can J Ophthalmol. 2020 Aug;55(4):293-300. doi: 10.1016/j.jcjo.2019.12.007. Epub 2020 Apr 3.

Abstract

OBJECTIVE

To explore segmentation errors, image quality, and motion-associated artifacts in eyes with idiopathic epiretinal membrane (ERM).

METHODS

This is a prospective observational study. We included 39 eyes affected by ERM and 40 eyes from age-matched healthy subjects. Optical coherence tomography-angiography (OCT-A) was performed in both groups. Segmentation was automatically performed by intergraded software. Segmentation was regarded as inaccurate if either border deviated from the correct plane by more than 50 μm. Presence of motion artifacts (blink lines, displacement, stretch artifacts, quilting, vessel doubling) and image quality index were reported.

RESULTS

Quality index score was 7.2 ± 0.9 for the ERM patients. Phakic eyes with ERM had quality index score of 7.71 ± 1.06, and pseudo-phakic eyes with ERM had a quality index score of 7.32 ± 0.85 (p = 0.22). Motion artifacts were 1.22 ± 0.7 in the study cohort. Segmentation was accurate in all healthy subjects (n = 40). Segmentation errors occurred in 64.1% of ERM patients. The inner plexiform layer was the segmentation boundary most prone to inaccurate segmentation, followed by the internal limiting membrane. Segmentation of retinal pigment epithelial layer was accurate in 96.7% of all cases.

CONCLUSIONS

OCT-A image quality cannot be accurately reproduced in pathological conditions, such as in ERM patients, and is prone to motion artifacts and segmentation errors. Incorrect segmentation results in anatomically incorrect en-face OCT-A images and subsequently in false quantification measures.

摘要

目的

探讨特发性视网膜内界膜(ERM)患者的分割错误、图像质量和运动相关伪影。

方法

这是一项前瞻性观察研究。我们纳入了 39 只受 ERM 影响的眼睛和 40 只年龄匹配的健康对照眼。两组均行光学相干断层扫描血管造影(OCT-A)检查。通过集成软件自动进行分割。如果边界偏离正确平面超过 50 μm,则认为分割不准确。报告存在运动伪影(眨眼线、位移、拉伸伪影、绗缝、血管重影)和图像质量指数。

结果

ERM 患者的质量指数评分为 7.2±0.9。有ERM 的 39 只眼,质量指数评分为 7.71±1.06,无ERM 的 39 只眼,质量指数评分为 7.32±0.85(p=0.22)。研究队列中运动伪影为 1.22±0.7。所有健康受试者(n=40)的分割均准确。ERM 患者中,64.1%的分割存在错误。最容易出现不准确分割的是内丛状层,其次是内界膜。所有病例的视网膜色素上皮层分割均准确,达 96.7%。

结论

OCT-A 图像质量在病理条件下(如 ERM 患者)无法准确再现,且容易出现运动伪影和分割错误。错误的分割会导致解剖上不正确的 OCT-A 图像,并随后导致错误的定量测量。