Suppr超能文献

新型旋转扩张器鞘管TightRail™用于导线拔除:双中心经验

Use of the new rotating dilator sheath TightRail™ for lead extraction: A bicentric experience.

作者信息

Mazzone Patrizio, Melillo Francesco, Radinovic Andrea, Marzi Alessandra, Paglino Gabriele, Della Bella Paolo, Mascioli Giosuè

机构信息

Arrhythmia Unit and Electrophysiology Laboratories Department of Cardiology and Cardiothoracic Surgery Ospedale San Raffaele Milan Italy.

Electrophysiology Unit Cardiovascular Department Humanitas Gavazzeni Hospital Bergamo Italy.

出版信息

J Arrhythm. 2020 Feb 14;36(2):343-350. doi: 10.1002/joa3.12310. eCollection 2020 Apr.

Abstract

AIM

The aim of this study is to assess the safety and efficacy of the new bidirectional rotational mechanical sheath TightRail™ (Spectranetics) for lead extraction.

METHODS AND RESULTS

This is a bicentric prospective study that included patients who underwent a transvenous lead extraction (TLE) in two Italian centers (San Raffaele Hospital and Humanitas Gavazzeni Hospital). From November 2016 to December 2018, 26 patients underwent a TLE procedure in which the TightRail™ was used. The new TightRail Sub-C was used in 20 (76%) patients to overcome the fibrosis between the vessel and the first rib. Median age was 69 (IQR 60.7-79.5) years. The indication for TLE were infection (57.7%) or lead dysfunction (42.3%). A total of 57 leads (range 1-4), 40 of which using the TightRail (range 1-4), were extracted. Overall mean implant duration was 98.2.0 ± 66.5 months. Mean age of the lead extracted with the TightRail sheath was 99.1 ± 70.2 months and was higher compared to that of the leads extracted manually (84.4 ± 60.3 months,  = .001). The overall clinical success was 100% and complete procedural success without the use of a snare was achieved in 98.3%. There were no cases of death or major complications and only two minor complications occurred. All patients were event-free at 6-month follow-up.

CONCLUSION

This initial experience using the TightRail™ suggests a high safety and efficacy profile for extractions in a wide range of lead age.

摘要

目的

本研究旨在评估新型双向旋转机械鞘管TightRail™(斯贝克特拉尼克斯公司)用于导线拔除的安全性和有效性。

方法与结果

这是一项双中心前瞻性研究,纳入了在意大利两个中心(圣拉斐尔医院和加瓦泽尼人文医院)接受经静脉导线拔除(TLE)的患者。2016年11月至2018年12月,26例患者接受了使用TightRail™的TLE手术。20例(76%)患者使用了新型TightRail Sub-C以克服血管与第一肋骨之间的纤维化。中位年龄为69岁(四分位间距60.7 - 79.5岁)。TLE的适应证为感染(57.7%)或导线功能障碍(42.3%)。共拔除57根导线(范围1 - 4根),其中40根使用TightRail(范围1 - 4根)。总体平均植入时间为98.2±66.5个月。使用TightRail鞘管拔除的导线平均使用年限为99.1±70.2个月,高于手动拔除的导线(84.4±60.3个月,P = 0.001)。总体临床成功率为100%,98.3%的患者在未使用圈套器的情况下实现了完全手术成功。无死亡或重大并发症病例,仅发生两例轻微并发症。所有患者在6个月随访时均无事件发生。

结论

使用TightRail™的这一初步经验表明,在广泛的导线使用年限范围内进行拔除时,其具有较高的安全性和有效性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/c65c/7132215/15e854b5f7d6/JOA3-36-343-g001.jpg

相似文献

1
Use of the new rotating dilator sheath TightRail™ for lead extraction: A bicentric experience.
J Arrhythm. 2020 Feb 14;36(2):343-350. doi: 10.1002/joa3.12310. eCollection 2020 Apr.
2
Initial experience with the TightRail™ Rotating Mechanical Dilator Sheath for transvenous lead extraction.
Europace. 2016 Jul;18(7):1043-8. doi: 10.1093/europace/euv245. Epub 2015 Oct 14.
3
Outcomes of transvenous lead extraction using the TightRail™ mechanical rotating dilator sheath and excimer laser sheath.
J Cardiovasc Electrophysiol. 2021 Jul;32(7):1969-1978. doi: 10.1111/jce.15105. Epub 2021 Jun 1.
4
Comparison of acute and long-term outcomes of Evolution and TightRail™ mechanical dilator sheaths during transvenous lead extraction.
J Cardiovasc Electrophysiol. 2021 May;32(5):1395-1404. doi: 10.1111/jce.15006. Epub 2021 Mar 25.
5
Efficacy and safety of the new TightRail™ mechanical sheath for transvenous lead extraction: Results of a French multicenter study.
J Cardiovasc Electrophysiol. 2022 Apr;33(4):731-737. doi: 10.1111/jce.15399. Epub 2022 Feb 11.
6
Comparison between TightRail rotating dilator sheath and GlideLight laser sheath for transvenous lead extraction.
Pacing Clin Electrophysiol. 2021 May;44(5):895-902. doi: 10.1111/pace.14206. Epub 2021 Mar 27.
7
Single center experience with the novel rotating dilator sheath TightRail Sub-C in transvenous lead extraction.
J Cardiovasc Electrophysiol. 2023 Aug;34(8):1738-1743. doi: 10.1111/jce.15970. Epub 2023 Jun 21.
8
Comparison of two types of rotational mechanical dilatator sheath: Evolution and TightRail.
Pacing Clin Electrophysiol. 2019 Sep;42(9):1226-1235. doi: 10.1111/pace.13755. Epub 2019 Jul 15.
10
Performance and outcomes of transvenous rotational lead extraction: Results from a prospective, monitored, international clinical study.
Heart Rhythm O2. 2021 Mar 2;2(2):113-121. doi: 10.1016/j.hroo.2021.02.005. eCollection 2021 Apr.

引用本文的文献

1
Prevention and Risk Assessment of Cardiac Device Infections in Clinical Practice.
J Clin Med. 2024 May 4;13(9):2707. doi: 10.3390/jcm13092707.
3
Transvenous Lead Extraction: Work in Progress.
Eur Cardiol. 2023 Jun 19;18:e44. doi: 10.15420/ecr.2023.06. eCollection 2023.
4
Practical Approaches to Transvenous Lead Extraction Procedures-Clinical Case Series.
Int J Environ Res Public Health. 2022 Dec 26;20(1):379. doi: 10.3390/ijerph20010379.
5
Working on the dirty side-the ipsilateral subclavian access for temporary pacing after lead extraction.
J Arrhythm. 2022 Feb 4;38(2):192-198. doi: 10.1002/joa3.12677. eCollection 2022 Apr.
7
Tricuspid insufficiency after cardiac-implantable electronic device placement.
J Community Hosp Intern Med Perspect. 2021 Nov 15;11(6):793-798. doi: 10.1080/20009666.2021.1967569. eCollection 2021.
8
Transvenous Lead Extraction without Procedure-Related Deaths in 1000 Consecutive Patients: A Single-Center Experience.
Vasc Health Risk Manag. 2021 Aug 5;17:445-459. doi: 10.2147/VHRM.S318205. eCollection 2021.
10
Safety and efficacy of transvenous lead extraction of very old leads.
Interact Cardiovasc Thorac Surg. 2021 Apr 8;32(3):402-407. doi: 10.1093/icvts/ivaa278.

本文引用的文献

2
Predictors of Intensive Care Unit Admission in Patients Undergoing Lead Extraction: A 10-Year Observational Study in a High-Volume Center.
J Cardiothorac Vasc Anesth. 2019 Jul;33(7):1845-1851. doi: 10.1053/j.jvca.2019.02.012. Epub 2019 Feb 13.
6
Initial experience with the TightRail™ Rotating Mechanical Dilator Sheath for transvenous lead extraction.
Europace. 2016 Jul;18(7):1043-8. doi: 10.1093/europace/euv245. Epub 2015 Oct 14.
7
Advanced techniques for chronic lead extraction: heading from the laser towards the evolution system.
Europace. 2013 Dec;15(12):1771-6. doi: 10.1093/europace/eut126. Epub 2013 May 3.
8
Transvenous lead extractions: comparison of laser vs. mechanical approach.
Europace. 2013 Nov;15(11):1636-41. doi: 10.1093/europace/eut086. Epub 2013 Apr 12.
10
Lead extraction is preferred for lead revisions and system upgrades: when less is more.
Circ Arrhythm Electrophysiol. 2010 Aug;3(4):413-24; discussion 424. doi: 10.1161/CIRCEP.110.954107.