定量微弹性成像在保乳手术切缘评估中的诊断准确性。

Diagnostic Accuracy of Quantitative Micro-Elastography for Margin Assessment in Breast-Conserving Surgery.

机构信息

BRITElab, Harry Perkins Institute of Medical Research, QEII Medical Centre, Nedlands, and Centre for Medical Research, The University of Western Australia, Perth, Australia.

School of Medicine, The University of Western Australia, Perth, Australia.

出版信息

Cancer Res. 2020 Apr 15;80(8):1773-1783. doi: 10.1158/0008-5472.CAN-19-1240.

Abstract

Inadequate margins in breast-conserving surgery (BCS) are associated with an increased likelihood of local recurrence of breast cancer. Currently, approximately 20% of BCS patients require repeat surgery due to inadequate margins at the initial operation. Implementation of an accurate, intraoperative margin assessment tool may reduce this re-excision rate. This study determined, for the first time, the diagnostic accuracy of quantitative micro-elastography (QME), an optical coherence tomography (OCT)-based elastography technique that produces images of tissue microscale elasticity, for detecting tumor within 1 mm of the margins of BCS specimens. Simultaneous OCT and QME were performed on the margins of intact, freshly excised specimens from 83 patients undergoing BCS and on dissected specimens from 7 patients undergoing mastectomy. The resulting three-dimensional images (45 × 45 × 1 mm) were coregistered with postoperative histology to determine tissue types present in each scan. Data from 12 BCS patients and the 7 mastectomy patients served to build a set of images for reader training. One hundred and fifty-four subimages (10 × 10 × 1 mm) from the remaining 71 BCS patients were included in a blinded reader study, which resulted in 69.0% sensitivity and 79.0% specificity using OCT images, versus 92.9% sensitivity and 96.4% specificity using elasticity images. The quantitative nature of QME also facilitated development of an automated reader, which resulted in 100.0% sensitivity and 97.7% specificity. These results demonstrate high accuracy of QME for detecting tumor within 1 mm of the margin and the potential for this technique to improve outcomes in BCS. SIGNIFICANCE: An optical imaging technology probes breast tissue elasticity to provide accurate assessment of tumor margin involvement in breast-conserving surgery.

摘要

保乳手术(BCS)中的不充分切缘与乳腺癌局部复发的可能性增加有关。目前,大约有 20%的 BCS 患者由于初始手术时切缘不充分,需要再次手术。实施一种准确的术中切缘评估工具可能会降低这种再次切除率。本研究首次确定了定量微弹性成像(QME)的诊断准确性,这是一种基于光相干断层扫描(OCT)的弹性成像技术,可生成组织微尺度弹性的图像,用于检测 BCS 标本切缘 1mm 内的肿瘤。对 83 例接受 BCS 手术的完整、新鲜切除标本的切缘以及 7 例接受乳房切除术的标本进行了同时 OCT 和 QME 检查。所得的三维图像(45×45×1mm)与术后组织学进行了配准,以确定每个扫描中存在的组织类型。来自 12 例 BCS 患者和 7 例乳房切除术患者的数据用于建立一组用于读者培训的图像。从其余 71 例 BCS 患者中选取了 154 个亚图像(10×10×1mm)进行了盲法读者研究,使用 OCT 图像的结果为 69.0%的灵敏度和 79.0%的特异性,而使用弹性图像的结果为 92.9%的灵敏度和 96.4%的特异性。QME 的定量性质还促进了自动读者的开发,其结果为 100.0%的灵敏度和 97.7%的特异性。这些结果表明,QME 对检测 1mm 内的肿瘤具有很高的准确性,并且该技术有可能改善 BCS 的结果。意义:一种光学成像技术探测乳腺组织的弹性,以提供对保乳手术中肿瘤切缘受累的准确评估。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验