Suppr超能文献

皮下植入式心律转复除颤器在左心室辅助装置患者中的故障排除:病例系列和系统评价。

Subcutaneous implantable cardioverter-defibrillator troubleshooting in patients with a left ventricular assist device: A case series and systematic review.

机构信息

Cardiac Electrophysiology Section, Duke Heart Center, Duke University Medical Center, Durham, North Carolina.

Cardiac Electrophysiology Section, Duke Heart Center, Duke University Medical Center, Durham, North Carolina.

出版信息

Heart Rhythm. 2020 Sep;17(9):1536-1544. doi: 10.1016/j.hrthm.2020.04.019. Epub 2020 Apr 15.

Abstract

BACKGROUND

There are limited data on the performance of the subcutaneous implantable cardioverter-defibrillator (S-ICD) in patients with a left ventricular assist device (LVAD).

OBJECTIVE

The purpose of this study was to describe the clinical course and outcomes of patients with both an S-ICD and an LVAD at our institution and via a systematic review of published studies.

METHODS

We performed a retrospective cohort study of all patients who underwent LVAD implantation from 2009 to 2019 at Duke University Hospital. We also performed a systematic review of studies involving patients with an S-ICD and LVAD using the PubMed/Embase databases.

RESULTS

Of 588 patients undergoing LVAD implantation with a preexisting implantable cardioverter-defibrillator, 4 had an S-ICD in situ after LVAD implantation. All 4 patients developed electromagnetic interference (EMI) in the primary/secondary vectors after LVAD implantation, resulting in inappropriate implantable cardioverter-defibrillator shocks in 2 patients. Sensing in the alternate vector was adequate immediately postoperatively in 1 patient. Postoperative undersensing was present in the alternate vector in 3 patients but improved at first outpatient follow-up in 2 patients, allowing tachy therapies to be reenabled. Eight studies involving 27 patients were identified in the systematic review. EMI was common and frequently absent in the alternate vector (6 of 7 patients).

CONCLUSION

Undersensing and EMI are common after LVAD implantation in patients with an S-ICD in situ, particularly in the primary and secondary sensing vectors. Undersensing in the alternate vector may improve during follow-up, obviating the need for device revision or extraction.

摘要

背景

目前关于皮下植入式心律转复除颤器(S-ICD)在左心室辅助装置(LVAD)患者中的应用效果的数据有限。

目的

本研究旨在通过对本机构和已发表研究的系统回顾,描述同时植入 S-ICD 和 LVAD 的患者的临床过程和结局。

方法

我们对 2009 年至 2019 年期间在杜克大学医院接受 LVAD 植入的所有患者进行了回顾性队列研究。我们还使用 PubMed/Embase 数据库对涉及 S-ICD 和 LVAD 患者的研究进行了系统回顾。

结果

在 588 例因预先存在的植入式心律转复除颤器而行 LVAD 植入的患者中,有 4 例在 LVAD 植入后存在原位 S-ICD。所有 4 例患者在 LVAD 植入后均出现初级/次级向量中的电磁干扰(EMI),导致 2 例患者出现不适当的植入式心律转复除颤器电击。1 例患者在术后即刻的备用向量中感知良好。3 例患者在备用向量中存在术后欠感知,但在 2 例患者的首次门诊随访中得到改善,从而使心动过速治疗得以重新启用。在系统回顾中,共确定了 8 项涉及 27 例患者的研究。EMI 很常见,并且常在备用向量中缺失(7 例患者中有 6 例)。

结论

在原位 S-ICD 的 LVAD 植入患者中,EMI 和欠感知很常见,特别是在初级和次级感知向量中。备用向量中的欠感知可能会在随访期间得到改善,从而避免了设备修订或提取的需要。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验