Jago J R, Murray A
Regional Medical Physics Department, Freeman Hospital, Newcastle upon Tyne, UK.
Clin Phys Physiol Meas. 1988 Nov;9(4):319-30. doi: 10.1088/0143-0815/9/4/003.
Peripheral pulses have been recorded and analysed to determine the accuracy with which pulse transit times (PTTs) can be measured. Measurements of PTT between the ECG Q-wave and various peripheral sites were made in 10 normal subjects on 10 separate days. Mean values were determined for the ears (174 ms), fingers (245 ms), and toes (361 ms). The technique was sufficiently accurate to detect small changes in PTT due to changes in posture; sitting to lying, 5.2 ms. When comparing simultaneous measurements on bilateral sites only small differences in PTT were discovered, and these were not significant in the study group as a whole. However, these differences were significant in some individuals. When the subjects raised a single arm or leg, significant differences (38 ms and 49 ms respectively) were recorded between sides. The day-to-day repeatability sigma (expressed as the square root of the within-subject mean square variance) of individual PTT measurements on a subject (supine) was for ears, fingers and toes respectively 9.4, 9.2 and 12 ms. For right-left differences the repeatability was 7.2, 5.9 and 14 ms. Hence changes in PTTs, or differences between right and left sides, can be detected from single measurements with 95% confidence if they exceed approximately 20 ms in ears or fingers and 30 ms in toes.
已记录并分析外周脉搏,以确定测量脉搏传输时间(PTT)的准确性。在10名正常受试者的10个不同日期,测量了心电图Q波与各个外周部位之间的PTT。确定了耳部(174毫秒)、手指(245毫秒)和脚趾(361毫秒)的平均值。该技术足够精确,能够检测因姿势变化导致的PTT微小变化;从坐姿变为躺姿时,变化为5.2毫秒。在比较双侧部位的同步测量时,仅发现PTT存在微小差异,且在整个研究组中这些差异不显著。然而,在一些个体中这些差异是显著的。当受试者抬起单臂或单腿时,两侧之间记录到显著差异(分别为38毫秒和49毫秒)。一名受试者(仰卧位)的个体PTT测量的每日重复性标准差(表示为受试者内均方方差的平方根),耳部、手指和脚趾分别为9.4、9.2和12毫秒。左右差异的重复性为7.2、5.9和14毫秒。因此,如果耳部或手指的PTT变化或左右两侧差异超过约20毫秒,脚趾超过30毫秒,那么单次测量就能以95%的置信度检测到这些变化。