Sharp Jennifer, Bouton Daniel, Anabell Lucas, Degan Timothy, Sienko Susan, Welborn Michelle C
Good Samaritan Hospital, Los Angeles, CA.
Shriners Hospital for Children, Portland, OR.
J Pediatr Orthop. 2020 Nov/Dec;40(10):587-591. doi: 10.1097/BPO.0000000000001570.
Biplanar digital slot scanning technology has become the standard of care in the treatment of scoliosis. Yet, the amount of distortion and reproducibility of this type of imaging modality has yet to be fully investigated. In our paper "Image distortion in biplanar slot scanning: part 1 patient-specific factors" we found that there was potentially clinically impactful interimage distortion. The purpose of this study was to evaluate the degree to which this image distortion was secondary to the image acquisition process.
Four 25 mm radio-opaque markers were placed at C3, T1, T12, and L5 on a full-length skeleton model. The skeleton was imaged in 10 different positions within the scanner. Five posteroanterior and 5 lateral images were obtained in each position. Two orthopaedic attending physicians and 3 orthopaedic resident physicians measured the markers for a total of 3200 measurements. Intraclass correlation coefficients (ICCs) and 95% confidence intervals were used to examine image distortion.
Average marker size was 24.77, with a standard error of measurement of 0.00493. Image distortion and standard error of measurement accounted for ∼0.5% to 1.5% of total the measurement. Overall, there was good reliability and consistency when looking at markers in different views (ICC 0.790), planes, and locations within the image. Horizontal measurements were found to be more consistent and have better reliability (ICC 0.881) than vertical measurements (ICC 0.386). Position within the scanner had minimal impact on the accuracy of the measurements.
This study demonstrates that there is minimal error due to image acquisition and measurement when using a biplanar slot scanner. Biplanar slot scanning technology tended to underestimate the size of the marker; however, the least accurate measurements only erred by 1.5% from the true length. This indicates that unlike traditional radiographs the sources of error in biplanar slot scanning images are not due to parallax and are likely due to patient-specific factors and rather than the technology itself.
双平面数字缝隙扫描技术已成为脊柱侧弯治疗的标准护理手段。然而,这种成像方式的失真量和可重复性尚未得到充分研究。在我们的论文《双平面缝隙扫描中的图像失真:第1部分患者特异性因素》中,我们发现存在可能具有临床影响的图像间失真。本研究的目的是评估这种图像失真在多大程度上是由图像采集过程导致的。
在一个全长骨骼模型上,于C3、T1、T12和L5处放置了4个25毫米的不透射线标记物。在扫描仪内将骨骼置于10个不同位置进行成像。每个位置获取5张前后位和5张侧位图像。两名骨科主治医师和三名骨科住院医师对标记物进行测量,共计3200次测量。使用组内相关系数(ICC)和95%置信区间来检查图像失真情况。
标记物的平均尺寸为24.77,测量标准误差为0.00493。图像失真和测量标准误差占总测量值的约0.5%至1.5%。总体而言,在不同视图(ICC 0.790)、平面以及图像内位置观察标记物时,具有良好的可靠性和一致性。发现水平测量比垂直测量更一致且可靠性更好(ICC 0.881对比ICC 0.386)。扫描仪内的位置对测量准确性的影响极小。
本研究表明,使用双平面缝隙扫描仪时,由于图像采集和测量导致的误差极小。双平面缝隙扫描技术往往会低估标记物的尺寸;然而,最不准确的测量值与真实长度的误差仅为1.5%。这表明与传统X线片不同,双平面缝隙扫描图像中的误差来源并非视差,而可能是患者特异性因素,而非技术本身。