School of Chemistry and Molecular Engineering, East China Normal University, 500 Dongchuan Road, Shanghai, 200241, China.
Anal Bioanal Chem. 2020 Jun;412(15):3737-3743. doi: 10.1007/s00216-020-02625-5. Epub 2020 May 3.
Extracellular pH can indicate the variation in organelle function and cell state. It is important to measure extracellular pH (pHe) with a controllable distance. In this work, a potentiometric SECM dual-microelectrode was developed to monitor the pHe of MCF-7 cells under electrical stimulation. The distance between the dual-microelectrode and the cells was determined first with a gold microelectrode by recording the approaching curve, and the pH was determined using an open-circuit potential (OCP) technique with a polyaniline-modified Pt microelectrode. The pH microelectrode showed a response slope of 53.0 ± 0.4 mV/pH and good reversibility from pH 4 to pH 8, fast response within 10 s, and a potential drift of 1.13% for 3 h, and thus was employed to monitor the pHe of stimulated cells. The value of pHe decreased with the decrease in the distance to cells, likely due to the release of H. With an increase in the stimulation potential or time, the pHe value decreased, as the cell membrane became more permeable, which was verified by fluorescence staining of calcein-AM/PI (propidium iodide). Based on these results, this method can be widely applied for determining the species released by biosystems at a controllable position.
细胞外 pH 值可以反映细胞器功能和细胞状态的变化。因此,需要使用可控制距离的方法来测量细胞外 pH 值(pHe)。在这项工作中,我们开发了一种基于 SECM 的双微电极电位法来监测 MCF-7 细胞在电刺激下的 pHe 值。首先,使用金微电极记录接近曲线来确定双微电极与细胞之间的距离,然后使用聚邻苯二胺修饰的 Pt 微电极的开路电位(OCP)技术来确定 pH 值。pH 值微电极的响应斜率为 53.0 ± 0.4 mV/pH,在 pH 值为 4 到 8 之间具有良好的可逆性,10 s 内快速响应,3 h 内电位漂移为 1.13%,因此可用于监测刺激细胞的 pHe 值。随着与细胞距离的减小,pHe 值降低,这可能是由于 H 的释放所致。随着刺激电位或时间的增加,细胞膜的通透性增加,pHe 值降低,这可以通过 calcein-AM/PI(碘化丙啶)荧光染色得到验证。基于这些结果,该方法可广泛应用于确定在可控制位置由生物体系释放的物质种类。