Camacho-Sandoval Rosa, Jiménez-Uribe Alexis, Tenorio-Calvo Alejandra V, López-Morales Carlos A, Muñoz-García Leslie, Montes-Luna Alejandra, García-Xolalpa Héctor Leonardo, Velasco-Velázquez Marco, Pavón Lenin, Pérez-Tapia Sonia Mayra, Medina-Rivero Emilio
Unidad de Desarrollo e Investigación en Bioprocesos. Escuela Nacional de Ciencias Biológicas. Instituto Politécnico Nacional, Ciudad de México, Mexico.
Sartorius de México S.A. de C.V. Tepotzotlán, Estado de México, Mexico.
Biotechnol Rep (Amst). 2020 Apr 19;26:e00456. doi: 10.1016/j.btre.2020.e00456. eCollection 2020 Jun.
Technological advances allowed the development of high-throughput instruments such as IntelliCyt iQue Screener PLUS®. Here, we took advantage of this technology to transfer a previously validated cytotoxicity assay. The evaluated parameters were cell permeability, caspase activation and phosphatidyl serine exposure. The assay was accurate (r = 0.90), precise (%CV ≤ 18.90) and specific. These results showed that this technology is suitable to be used in control quality environments. In addition, the automation provided a faster acquisition and analysis of data with precise and accurate results. This application could be implemented to evaluate another mechanism of action of different biotherapeutics.
技术进步推动了诸如IntelliCyt iQue Screener PLUS®等高通量仪器的发展。在此,我们利用这项技术转移了先前经过验证的细胞毒性检测方法。评估的参数包括细胞通透性、半胱天冬酶激活和磷脂酰丝氨酸暴露。该检测方法准确(r = 0.90)、精密(%CV≤18.90)且具有特异性。这些结果表明,这项技术适用于质量控制环境。此外,自动化操作实现了更快的数据采集与分析,结果精确且准确。该应用可用于评估不同生物治疗药物的另一种作用机制。