Suppr超能文献

采用气相色谱/电子捕获负离子化学电离质谱联用技术对人血浆中的可乐定进行定量测定。

Quantitative measurement of clonidine in human plasma by combined gas chromatography/electron capture negative ion chemical ionization mass spectrometry.

作者信息

Girault J, Fourtillan J B

机构信息

Laboratoire de Parmacocinétique, Hopital Jean Bernard, Poitiers, France.

出版信息

Biomed Environ Mass Spectrom. 1988 Dec;17(6):443-8. doi: 10.1002/bms.1200170605.

Abstract

A new quantitative and selective assay was developed for the measurement of low levels of clonidine in human plasma. The drug and the deuterated internal standard (2H4)clonidine are quantified by gas chromatography/negative ion chemical ionization mass spectrometry with methane as the reagent gas. After a three-step extraction procedure, the two compounds of interest are converted to their fluorinated derivatives using 3,5-bis(trifluoromethyl)benzoyl chloride. The mass spectrometer is focused to monitor abundant and stable characteristic high-mass ions (m/z 673 and 709) which are generated in the ion source by a resonance electron capture process. This assay requires only 1 ml of plasma sample and the limit of detection of the method is 5 pg ml-1 with a relative standard deviation of less than 7%.

摘要

开发了一种新的定量和选择性测定方法,用于测量人血浆中低水平的可乐定。通过气相色谱/负离子化学电离质谱法,以甲烷作为反应气,对药物和氘代内标(2H4)可乐定进行定量。经过三步萃取程序后,使用3,5-双(三氟甲基)苯甲酰氯将两种目标化合物转化为它们的氟化衍生物。质谱仪聚焦于监测在离子源中通过共振电子捕获过程产生的丰富且稳定的特征高质量离子(m/z 673和709)。该测定方法仅需要1毫升血浆样本,方法的检测限为5皮克/毫升,相对标准偏差小于7%。