Commonly, a dysfunctional defibrillator lead is abandoned and a new lead is implanted. Long-term follow-up data on abandoned leads are sparse. We aimed to investigate the incidence and reasons for extraction of abandoned defibrillator leads in a nationwide cohort and to describe extraction procedure-related complications.
All abandoned transvenous defibrillator leads were identified in the Danish Pacemaker and ICD Register from 1991 to 2019. The event-free survival of abandoned defibrillator leads was studied, and medical records of patients with interventions on abandoned defibrillator leads were audited for procedure-related data. We identified 740 abandoned defibrillator leads. Meantime from implantation to abandonment was 7.2 ± 3.8 years with mean patient age at abandonment of 66.5 ± 13.7 years. During a mean follow-up after abandonment of 4.4 ± 3.1 years, 65 (8.8%) abandoned defibrillator leads were extracted. Most frequent reason for extraction was infection (pocket and systemic) in 41 (63%) patients. Procedural outcome after lead extraction was clinical success in 63 (97%) patients. Minor complications occurred in 3 (5%) patients, and major complications in 1 (2%) patient. No patient died from complication to the procedure during 30-day follow-up after extraction.
More than 90% of abandoned defibrillator leads do not need to be extracted during long-term follow-up. The most common indication for extraction is infection. Abandoned defibrillator leads can be extracted with high clinical success rate and low risk of major complications at high-volume centres.
通常情况下,功能失调的除颤器导联会被废弃,并植入新的导联。废弃导联的长期随访数据较为缺乏。本研究旨在调查全国性队列中废弃除颤器导联的拔除发生率和原因,并描述与拔除程序相关的并发症。
在 1991 年至 2019 年间,我们在丹麦起搏器和 ICD 登记处确定了所有废弃的经静脉除颤器导联。研究了废弃除颤器导联的无事件生存率,并对接受废弃除颤器导联干预的患者的病历进行了审核,以获取与程序相关的数据。我们共确定了 740 根废弃的除颤器导联。从植入到废弃的平均时间为 7.2±3.8 年,废弃时患者的平均年龄为 66.5±13.7 岁。在废弃后的平均随访时间 4.4±3.1 年内,有 65 根(8.8%)废弃的除颤器导联被拔除。最常见的拔除原因是感染(囊袋和全身),在 41 名(63%)患者中发生。63 名(97%)患者的拔除后临床结局为成功。3 名(5%)患者出现轻微并发症,1 名(2%)患者出现严重并发症。在拔除后的 30 天随访中,无患者因并发症而死亡。
在长期随访中,超过 90%的废弃除颤器导联不需要拔除。最常见的拔除指征是感染。在高容量中心,废弃的除颤器导联可以安全地拔除,具有较高的临床成功率和较低的严重并发症风险。