Br J Oral Maxillofac Surg. 2020 Oct;58(8):940-946. doi: 10.1016/j.bjoms.2020.04.034. Epub 2020 May 21.
The benefits of three-dimensional planning and guided surgery have been realised over the last few years in maxillofacial surgery. Reproducing the exact positioning of the cutting and drilling guides on the flat mandibular angles defined by the engineer is a challenge for the surgeon and for the reliability of guided bilateral sagittal split osteotomy. Reference screws positioned on the skeleton before the acquisition of medical computed tomographic data can provide a fixed landmark that can be used during surgery and by the engineer during the design phase. The objective of this proof of concept in vitro study is to calculate the accuracy obtained for guides positioned by inserting a reference screw. The precision obtained for 30 guides following the insertion of 30 reference screws on 15 mandibular models was analysed. The models were scanned using an optical scanner and compared to CAD-CAM projects. The mean (SD) absolute position (in)-inaccuracy is 0.1616 (0.1141) mm for the entire guide surface and 0.13143 (0.0835) mm for the rim surface. The results indicate that the use of reference screws is efficient, and so they can be used to position guides accurately during guided bilateral sagittal split osteotomy.
在过去几年里,三维规划和导向手术的优势在颌面外科领域得以实现。对于外科医生而言,要在由工程师定义的平坦下颌角上精确重现切割和钻孔导向器的定位,以及保证导向双侧矢状劈开截骨术的可靠性,都是一项挑战。在获取医学计算机断层扫描数据之前,将参考螺钉置于骨骼上,可为手术过程以及工程师在设计阶段提供一个固定的标志点。这项体外概念验证研究的目的是计算通过插入参考螺钉来定位导向器所获得的精度。分析了在15个下颌模型上插入30颗参考螺钉后,30个导向器所获得的精度。使用光学扫描仪对模型进行扫描,并与计算机辅助设计与制造(CAD-CAM)方案进行比较。整个导向器表面的平均(标准差)绝对位置误差为0.1616(0.1141)毫米,边缘表面为0.13143(0.0835)毫米。结果表明,使用参考螺钉是有效的,因此它们可用于在导向双侧矢状劈开截骨术中精确地定位导向器。