一种普通抛光锥形股骨柄的植入物骨折发生率较高。

High rates of implant fracture of a generic polished tapered femoral stem.

作者信息

Vanbiervliet Jens, Braem Annabel, Simon Jean-Pierre, Van Humbeek Jan, Brouwers Jonas, Ghijselings Stijn

机构信息

Orthopaedic Department, University Hospital Leuven, Leuven, Belgium.

Department of Materials Engineering, Catholic University Leuven, Leuven, Belgium.

出版信息

Hip Int. 2022 Jan;32(1):99-105. doi: 10.1177/1120700020925739. Epub 2020 May 27.

Abstract

OBJECTIVES

Cemented polished tapered stems have demonstrated excellent long-term outcomes. Based on this concept, many generic tapered stems have been released into the market. The aim of this study was to evaluate implant-related complications of 1 specific stem design.

METHODS

Between 2010 and 2017, 315 total hip replacements were performed using a Fortress stem (Biotechni, La Ciotat, France). Patient records and radiology were retrospectively reviewed for implant-related complications. A failure analysis was performed on the failed Fortress stems in order to determine the cause of premature failure.

RESULTS

7 (2.2%) patients sustained a fracture of the neck of the implant after a mean of 5 years (range 50-81 months). All fractures were atraumatic, originating at the introducer inlet of the stem. All fractured occurred in obese patients (BMI >33 kg/m) with a small sized prosthesis. Of these, there were 5 135° and 2 125° stems. Fracture risk was 23% (7/30) for patients with a small sized stem and a BMI >30 kg/m. All cases were revised using a cement-in-cement technique or a cementless modular revision stem. Failure analysis on the retrieved stems revealed a stress riser at the bottom of the introducer inlet.

CONCLUSIONS

An alarmingly high rate of early implant fractures was seen using this specific type of cemented stem, particularly when using smaller implant sizes in obese patients. Although based on a proven design, a specific modification led to a stress riser in the neck area, which resulted in a high incidence of implant failure. This series underlines the importance of a stepwise introduction into the market of new orthopaedic devices even when based on established concepts. Generic stems may not behave as the original stem upon which it was designed.

摘要

目的

骨水泥固定的抛光锥形柄已显示出优异的长期效果。基于这一理念,许多通用锥形柄已投放市场。本研究的目的是评估一种特定柄设计的植入物相关并发症。

方法

2010年至2017年期间,使用Fortress柄(法国拉西奥塔的Biotechni公司)进行了315例全髋关节置换术。对患者记录和影像学资料进行回顾性分析,以确定植入物相关并发症。对失败的Fortress柄进行失效分析,以确定过早失败的原因。

结果

7例(2.2%)患者在平均5年(范围50 - 81个月)后发生植入物颈部骨折。所有骨折均为非创伤性,起源于柄的插入器入口处。所有骨折均发生在肥胖患者(BMI>33 kg/m²)且使用小号假体的情况下。其中,有5个135°柄和2个125°柄。对于使用小号柄且BMI>30 kg/m²的患者,骨折风险为23%(7/30)。所有病例均采用骨水泥加骨水泥技术或非骨水泥模块化翻修柄进行翻修。对取出的柄进行的失效分析显示,插入器入口底部存在应力集中。

结论

使用这种特定类型的骨水泥柄时,早期植入物骨折的发生率高得惊人,尤其是在肥胖患者中使用较小尺寸的植入物时。尽管基于经过验证的设计,但特定的修改导致颈部区域出现应力集中,从而导致植入物失败的发生率很高。本系列研究强调了即使基于既定概念,新骨科器械逐步推向市场的重要性。通用柄的表现可能与设计它所基于的原始柄不同。