Pickup J C, Claremont D J
Division of Chemical Pathology, United Medical School, Guy's Hospital, London, United Kingdom.
Horm Metab Res Suppl. 1988;20:34-6.
A range of miniature, amperometric enzyme electrodes have been constructed using the principle of mediated transfer of electrons. Ferrocene, or one of its derivatives, can be used in conjunction with immobilised glucose oxidase to produce sensors which are linear to at least 20 mmol/l glucose, have fast response times, are relatively oxygen insensitive and are suitable for implantation studies. In preliminary experiments in non-diabetic pigs, electrode responses followed blood responses, although mean tissue levels were about one fifth of those in blood. The ferrocene technology thus appears to be suitable for adaptation to the construction of implantable glucose electrodes for use in diabetic man.
利用电子介导转移原理构建了一系列微型安培酶电极。二茂铁或其衍生物之一可与固定化葡萄糖氧化酶结合使用,以生产对至少20毫摩尔/升葡萄糖呈线性响应、响应时间快、对氧气相对不敏感且适用于植入研究的传感器。在非糖尿病猪的初步实验中,电极反应跟随血液反应,尽管平均组织水平约为血液中的五分之一。因此,二茂铁技术似乎适合用于构建供糖尿病患者使用的植入式葡萄糖电极。