Kerner W, Zier H, Steinbach G, Brückel J, Pfeiffer E F, Weiss T, Cammann K, Planck H
Abteilung für Innere Medizin I, Universität Ulm.
Horm Metab Res Suppl. 1988;20:8-13.
A membrane limited amperometric enzyme electrode suitable for glucose measurement in biological fluids was developed. The sensor consists of a central platinum wire (0.3 mm) surrounded by a stainless steel tubing (0.8-1.0 mm outer diameter; 2-4 cm length). By successive dipcoating procedures, layers from cellulose acetate, glucose oxidase (crosslinked with glutaraldehyde) and polyurethane are placed on its surface. The platinum is polarized at +700 mV against steel. In vitro results: Electrodes are stable for at least 6 days. They exhibit a linear range extending to 500 mg/dl glucose. Response times are less than 100 sec. The sensors are not dependent on stirring and are relatively insensitive to changes of pH. Dependency of glucose measurement upon dissolved oxygen is negligible at oxygen concentrations above 0.5 mg/l. In vivo results: Preliminary studies in sheep using subcutaneously implanted needles indicate that short term glucose monitoring is feasible.
开发了一种适用于生物流体中葡萄糖测量的膜限电流型酶电极。该传感器由一根中心铂丝(0.3毫米)组成,周围是一根不锈钢管(外径0.8 - 1.0毫米;长度2 - 4厘米)。通过连续浸涂工艺,将醋酸纤维素、葡萄糖氧化酶(用戊二醛交联)和聚氨酯层置于其表面。铂相对于钢在+700毫伏下极化。体外结果:电极至少稳定6天。它们的线性范围扩展到500毫克/分升葡萄糖。响应时间小于100秒。传感器不依赖搅拌,对pH变化相对不敏感。在氧浓度高于0.5毫克/升时,葡萄糖测量对溶解氧的依赖性可忽略不计。体内结果:在绵羊身上使用皮下植入针进行的初步研究表明,短期葡萄糖监测是可行的。