评价 VIASURE 性传播疾病实时 PCR 检测试剂盒(CerTest Biotec)在性传播感染诊断中的应用。

Evaluation of VIASURE Sexually Transmitted Diseases Real Time PCR Detection kit (CerTest Biotec) for the diagnostic of sexually transmitted infections.

机构信息

Servicio de Microbiología, Laboratorio Clínico Metropolitana Norte, Hospital Universitari Germans Trias i Pujol, Badalona, Spain.

Servicio de Microbiología, Laboratorio Clínico Metropolitana Norte, Hospital Universitari Germans Trias i Pujol, Badalona, Spain.

出版信息

Enferm Infecc Microbiol Clin (Engl Ed). 2021 May;39(5):229-233. doi: 10.1016/j.eimc.2020.04.013. Epub 2020 Jun 1.

Abstract

INTRODUCTION

Sexually transmitted infections (STIs) are common in our environment, and trends have been increasing in the last few years. Different methods for STIs diagnosis have been applied by microbiology laboratories over years, but real-time PCR has improved this process. Our objective was to evaluate VIASURE Sexually Transmitted Diseases Real-Time PCR Detection kit (CerTestBiotec, S.L.) comparing with the real-time PCR technique used in our laboratory (Allplex™ STI7 Essential Assay, Seegene) which was considered as reference assay.

METHODS

A total of 948 samples from different sites (vaginal, endocervical, urethral, rectal, pharyngeal swabs and urine samples) were analyzed from July to September 2018.

RESULTS

A discordant result was obtained in 4.5% (43 samples). These discrepancies were mainly observed in threshold cycle (Ct) value next to the limit of detection. The k coefficient obtained shows a very high agreement between both methods with k values from 0.92 to 0.99.

CONCLUSIONS

VIASURE Sexually Transmitted Diseases Real-Time PCR Detection kit provides a very good correlation with Allplex STI7 and therefore, it's a good tool for the diagnostic of STIs. Positive results with Ct value obtained from 35 and low amplification signal should be applied with caution and should be interpreted based on the patient's clinical data.

摘要

简介

性传播感染(STIs)在我们的环境中很常见,近年来呈上升趋势。多年来,微生物学实验室已经应用了不同的 STI 诊断方法,但实时 PCR 已经改进了这一过程。我们的目的是评估 VIASURE 性传播疾病实时 PCR 检测试剂盒(CerTestBiotec,S.L.)与我们实验室使用的实时 PCR 技术(Allplex™ STI7 基本检测试剂盒,Seegene)进行比较,后者被认为是参考检测方法。

方法

共分析了 2018 年 7 月至 9 月间来自不同部位(阴道、宫颈、尿道、直肠、咽拭子和尿液样本)的 948 个样本。

结果

4.5%(43 个样本)得到了不一致的结果。这些差异主要发生在接近检测限的阈值循环(Ct)值附近。获得的 k 系数表明两种方法之间具有非常高的一致性,k 值从 0.92 到 0.99。

结论

VIASURE 性传播疾病实时 PCR 检测试剂盒与 Allplex STI7 具有很好的相关性,因此是诊断 STIs 的良好工具。Ct 值为 35 且扩增信号低的阳性结果应谨慎应用,并应根据患者的临床数据进行解释。