Suppr超能文献

使用库尔特S-Plus IV/D型血细胞计数器检测冷球蛋白

Detection of cryoglobulins by Coulter Counter model S-Plus IV/D.

作者信息

Banfi G, Bonini P A

机构信息

Laboratorio Analisi, Istituto Scientifico S. Raffaele, Milan, Italy.

出版信息

Clin Lab Haematol. 1988;10(4):453-9. doi: 10.1111/j.1365-2257.1988.tb01194.x.

Abstract

Cryoglobulins may produce interference in results from electronic particle counters. Depending on their size, cold-precipitable globulin crystals can be counted as leucocytes and/or platelets. The Coulter Counter model S-Plus IV/D permits detection of this interference and its associated spurious counts by producing a characteristic WBC histogram accompanied by a leucocyte 'flag'. Using this instrument over a 1-year period, 10 cases of cryoglobulinaemia were detected, four from patients with lymphoproliferative disease and one from a patient with autoimmune disease. The remaining five were classified as essential mixed cryoglobulinaemia. The reproducibility of this instrument anomaly permits identification of the presence of cryoglobulins.

摘要

冷球蛋白可能会干扰电子粒子计数器的检测结果。根据其大小,冷沉淀球蛋白晶体可被计为白细胞和/或血小板。库尔特S-Plus IV/D型血细胞计数器通过生成特征性的白细胞直方图并伴有白细胞“标记”,能够检测到这种干扰及其相关的假性计数。在1年的时间里使用该仪器,共检测到10例冷球蛋白血症患者,其中4例来自淋巴增殖性疾病患者,1例来自自身免疫性疾病患者。其余5例被归类为原发性混合性冷球蛋白血症。该仪器异常情况的可重复性有助于识别冷球蛋白的存在。