Ultrasound Med Biol. 2020 Sep;46(9):2265-2275. doi: 10.1016/j.ultrasmedbio.2020.05.005. Epub 2020 Jun 14.
The capillary dialyzer represents the central element of the extracorporeal blood circuit of a therapy system for hemodialysis. The aim of this study was to assess the blood-flow characteristics of dialyzers with the help of modern ultrasound techniques. Five brand-new dialyzers (FX80 classix, Fresenius Medical Care, Bad Homburg, Germany) and five dialyzers after a dialysis session were analyzed by different ultrasound techniques to detect functional and structural changes. B-mode and Doppler techniques were not suitable to describe differences in brand-new and clinically applied dialyzers. Contrast-enhanced ultrasonography, however, was able to visualize blood-flow profiles in the capillaries. Although dialyzers displayed no signs of clinical dysfunction, contrast-enhanced ultrasonography was able to detect blocked capillaries of varying degrees after a dialysis session in all five examined dialyzers. Consequently, the blood-flow velocity was higher in the remaining unblocked capillaries in comparison to the velocity in the brand-new dialyzers. This information may be helpful for improving the geometric design of dialyzers, including their capillary membranes, and optimizing anti-coagulation strategies in hemodialysis patients.
毛细血管透析器是血液透析治疗系统体外血液回路的核心元件。本研究旨在借助现代超声技术评估透析器的血流特性。我们使用不同的超声技术分析了 5 个全新的透析器(FX80 classix,费森尤斯医疗保健,德国巴特洪堡)和 5 个使用后的透析器,以检测功能和结构变化。B 型和多普勒技术不适合描述全新和临床应用的透析器之间的差异。然而,对比增强超声能够在毛细血管中可视化血流轮廓。尽管透析器没有显示出临床功能障碍的迹象,但在所有 5 个检查的透析器中,对比增强超声在透析后均能检测到不同程度的堵塞毛细血管。因此,与全新透析器相比,剩余未堵塞毛细血管中的血流速度更高。这些信息可能有助于改进透析器的几何设计,包括其毛细血管膜,并优化血液透析患者的抗凝策略。