Blood Press. 2020 Dec;29(6):375-381. doi: 10.1080/08037051.2020.1785273. Epub 2020 Jul 1.
Cuffless blood pressure (CL-BP) measurements are believed to be a potentially alternative to cuff-occlusion-based (C-BP) measurement. A new cuffless device was developed for ambulatory BP monitoring. We assessed the accuracy of a new CL-BP device compared to a standard oscillatory C-BP device over the 24 h.
Eighty-four consecutive patients were included in the study. BP was measured simultaneously by the CL-BP device and by a C-BP device over the 24 h. Calculations included 24 h mean systolic (S) BP, the mean diastolic (D) BP and the heart rate (HR). Correlations between the CL-BP and C-BP measurements were sought using Pearson's correlation coefficients and Bland-Altman plots.
Using the C-BP device, the 24 h SBP value for the cohort was 125.4 ± 10.9 mmHg (mean ± SD); the corresponding DBP value being 75 ± 8.3 mmHg. Mean SBP/DBP were higher with the CL-BP device, i.e. 131.1 ± 15.9/80.2 ± 9.7 mmHg . The correlation coefficients between the two sets of values were significant (SBP: = 0.58, DBP: = 0.65). Better correlations for SBP and DBP were found 1) in patients with BMI > 25 (SBP: = 0.65, DBP: = 0.70) compared to those with BMI <25 and 2) in males compared to females (SBP: = 0.71, DBP: = 0.77).
In our patients a CL-BP device estimated 24 h mean SBP and DBP differently from the classical oscillometric device, with a moderate correlation. CL-BP measurements were most accurate on male and overweight subjects.
无袖带血压(CL-BP)测量被认为是替代袖带加压法(C-BP)测量的一种潜在方法。一种新的无袖带设备被开发用于动态血压监测。我们评估了一种新的 CL-BP 设备与标准振荡式 C-BP 设备在 24 小时内的准确性。
84 名连续患者纳入研究。BP 通过 CL-BP 设备和 C-BP 设备同时测量 24 小时。计算包括 24 小时平均收缩压(S)BP、平均舒张压(D)BP 和心率(HR)。使用 Pearson 相关系数和 Bland-Altman 图寻找 CL-BP 和 C-BP 测量之间的相关性。
使用 C-BP 设备,队列的 24 小时 SBP 值为 125.4±10.9mmHg(平均值±标准差);相应的 DBP 值为 75±8.3mmHg。CL-BP 设备的平均 SBP/DBP 值更高,即 131.1±15.9/80.2±9.7mmHg。两组值之间的相关系数具有显著意义(SBP:=0.58,DBP:=0.65)。在 BMI>25 的患者中,SBP 和 DBP 的相关性更好(SBP:=0.65,DBP:=0.70),而在 BMI<25 的患者中,SBP 和 DBP 的相关性更好(SBP:=0.65,DBP:=0.70),而在 BMI<25 的患者中,SBP 和 DBP 的相关性更好(SBP:=0.65,DBP:=0.70),而在 BMI<25 的患者中,SBP 和 DBP 的相关性更好(SBP:=0.65,DBP:=0.70);与女性相比,男性的相关性更好(SBP:=0.71,DBP:=0.77)。
在我们的患者中,CL-BP 设备与经典的振荡式设备相比,估计 24 小时平均 SBP 和 DBP 不同,相关性中等。CL-BP 测量在男性和超重患者中最准确。