DeBerardinis Jessica, Neilsen Conner, Lidstone Daniel E, Dufek Janet S, Trabia Mohamed B
Department of Mechanical Engineering, University of Nevada Las Vegas, Las Vegas, USA.
Department of Kinesiology and Nutrition Sciences, University of Nevada Las Vegas, Las Vegas, USA.
J Rehabil Assist Technol Eng. 2020 Jul 6;7:2055668320921063. doi: 10.1177/2055668320921063. eCollection 2020 Jan-Dec.
Force platforms and pressure-measuring insoles are the most common tools used for measuring center of pressure. Earlier studies to assess these instruments suffered from limited sample sizes or an inadequate range of participant foot sizes. The purpose of this study was to propose new methods to extract and calculate comparably accurate center of pressure for the Kistler® force platform and Medilogic® insoles.
Center of pressure data were collected from 65 participants wearing pressure-measuring insoles (six different sizes). Participants walked over consecutive force platforms for three trials while wearing pressure-measuring insoles within socks. Onset force thresholds and center of pressure segment length thresholds were used to determine accurate center of pressure path length and width. A single step for each foot and trial was extracted from both instruments.
A strong correlation was observed between instruments in center of pressure length (4.12 ± 6.72% difference, r = 0.74). Center of pressure width varied and was weakly correlated (-7.04 ± 4.48% difference, r = 0.11).
The results indicate that both instruments can measure center of pressure path length consistently and with comparable accuracy (differences < 10%). There were differences between instruments in measuring center of pressure path width, which were attributed to the limited number of sensors across the width of the insoles.
测力平台和压力测量鞋垫是用于测量压力中心的最常用工具。早期评估这些仪器的研究存在样本量有限或参与者脚尺寸范围不足的问题。本研究的目的是提出新的方法,为奇石乐®测力平台和美迪龙®鞋垫提取并计算出相当准确的压力中心。
从65名穿着压力测量鞋垫(六种不同尺寸)的参与者收集压力中心数据。参与者穿着袜子内的压力测量鞋垫,在连续的测力平台上行走三次。起始力阈值和压力中心段长度阈值用于确定准确的压力中心路径长度和宽度。从两种仪器中提取每只脚每次试验的单步数据。
在压力中心长度方面,两种仪器之间观察到强相关性(差异为4.12±6.72%,r=0.74)。压力中心宽度存在差异且相关性较弱(差异为-7.04±4.48%,r=0.11)。
结果表明,两种仪器都能一致地测量压力中心路径长度,且精度相当(差异<10%)。两种仪器在测量压力中心路径宽度方面存在差异,这归因于鞋垫宽度方向上传感器数量有限。