Suppr超能文献

Anesthesia systems.

出版信息

Health Devices. 1988 Dec;17(12):366-7.

PMID:3273928
Abstract

We evaluated twelve disposable pressure transducers from eleven manufacturers. All were rated Acceptable and were ranked in four groups. The first two groups had similar performance characteristics. Groups three and four were ranked primarily according to the effect of fluid spills and the ease of achieving accurate dynamic response.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验