Cordier G
Pathol Biol (Paris). 1977 Apr;25(4):275-9.
Detection of gamma emitting isotopes is achieved conventionally by using a gamma spectrometer, or in some cases by liquid scintillation counting. We applied these two methods to the detection of 51 Chromium released in a lymphocyte-mediated cytotoxic reaction. Detection in liquid scintillation appeared more efficient than conventional gamma detection: 16.74% vs 1,11%. It is therefore possible to reduce by a factor of 10 the number of target cells in the cytotoxic assay if liquid scintillation counting is used. Either method gave the same results when applied on the same samples. Liquid scintillation counting is to be preferred when high sensitivity is needed. In addition it offers the advantage of using the same apparatus for detection of gamma or beta emitting isotopes.
传统上,通过使用伽马能谱仪来实现对发射伽马射线的同位素的检测,在某些情况下则通过液体闪烁计数法来检测。我们将这两种方法应用于检测淋巴细胞介导的细胞毒性反应中释放的51铬。液体闪烁检测似乎比传统的伽马检测更有效:分别为16.74%和1.11%。因此,如果使用液体闪烁计数法,细胞毒性试验中靶细胞的数量可以减少10倍。当对相同样本应用这两种方法时,得到的结果相同。当需要高灵敏度时,液体闪烁计数法更可取。此外,它还具有使用同一仪器检测发射伽马或贝塔射线的同位素的优点。