Department of Forensic Science, The George Washington University, Washington, DC, USA.
J Sep Sci. 2020 Oct;43(19):3756-3764. doi: 10.1002/jssc.202000767. Epub 2020 Aug 19.
This study ascertained the viability of a portable liquid chromatograph, operating at low microliter per minute flow, for the analysis of seized drugs at remote sites as well as in laboratory settings. Synthetic cathinones were screened using dual capillary columns in series, C8 and biphenyl, with on-column ultraviolet detection at 255 and 275 nm. The relative retention times of the two columns in series and their peak area absorbance ratio were used to determine if the 16 synthetic cathinones investigated could be uniquely identified in these conditions. Based on these parameters all of the analytes could be distinguished. Representative mixtures of synthetic cathinones were used to determine the repeatability, linearity, and limits of detection of the method. This cost effective and green instrumentation has the potential to satisfy minimum international guidelines for the analysis of seized drugs.