This study evaluated the calcium blooming-reducing effect and the differences of luminal diameter among various-energy virtual monochromatic images (VMIs) using rapid kilovolt-switching dual-energy computed tomography (DECT).
Forty-five calcified segments in 31 patients were analyzed. For the analysis, 40- to 140-keV VMIs on both non-contrast CT and coronary CT angiography were generated at 10-keV steps, and calcification size and luminal diameter were measured using CT number profile curve and full-width at half-maximum method. We compared calcification size and luminal diameter on each keV VMIs with those on 70-keV VMI.
There was no significant differences among the 40- to 140-keV VMIs regarding calcification size or luminal diameter.
The 40- to 140-keV VMIs produced by single-source DECT had no effect on the calcification size or luminal diameter in the coronary artery.
本研究采用快速千伏切换双能计算机断层扫描(DECT)评估了不同能量虚拟单色图像(VMIs)的钙辉现象减少效果及管腔直径差异。
对 31 例患者的 45 个钙化节段进行了分析。在分析中,采用 10keV 步长生成非对比 CT 和冠状动脉 CT 血管造影的 40-140keV VMIs,并使用 CT 值轮廓曲线和半最大值全宽法测量钙化大小和管腔直径。我们将每个 keV VMIs 上的钙化大小和管腔直径与 70keV VMI 上的值进行了比较。
在钙化大小或管腔直径方面,40-140keV VMIs 之间没有显著差异。
单源 DECT 生成的 40-140keV VMIs 对冠状动脉内的钙化大小或管腔直径没有影响。