The purpose of this study is to evaluate the effects of cone-beam computed tomography (CBCT) on dose distribution and normal tissue complication probability (NTCP) by constructing a comprehensive dose evaluation system for prostate intensity-modulated radiation therapy (IMRT).
A system that could combine CBCT and treatment doses with MATLAB was constructed. Twenty patients treated with prostate IMRT were studied. A mean dose of 78 Gy was prescribed to the prostate region, excluding the rectal volume from the target volume, with margins of 4 mm to the dorsal side of the prostate and 7 mm to the entire circumference. CBCT and treatment doses were combined, and the dose distribution and the NTCP of the rectum and bladder were evaluated.
The radiation dose delivered to 2% and 98% of the target volume increased by 0.90 and 0.74 Gy on average, respectively, in the half-fan mode and on average 0.76 and 0.72 Gy, respectively, in the full-fan mode. The homogeneity index remained constant. The percent volume of the rectum and bladder irradiated at each dose increased slightly, with a maximum increase of <1%. The rectal NTCP increased by approximately 0.07% from 0.46% to 0.53% with the addition of a CBCT dose, while the maximum NTCP in the bladder was approximately 0.02%.
This study demonstrated a method to evaluate a combined dose of CBCT and a treatment dose using the constructed system. The combined dose distribution revealed increases of <1% volume in the rectal and bladder doses and approximately 0.07% in the rectal NTCP.
本研究的目的是通过构建前列腺调强放射治疗(IMRT)的综合剂量评估系统,评估锥形束计算机断层扫描(CBCT)对剂量分布和正常组织并发症概率(NTCP)的影响。
构建了一个能将CBCT和治疗剂量与MATLAB相结合的系统。对20例接受前列腺IMRT治疗的患者进行了研究。前列腺区域的处方平均剂量为78 Gy,靶区体积不包括直肠体积,前列腺背侧边缘为4 mm,整个周长边缘为7 mm。将CBCT和治疗剂量相结合,评估直肠和膀胱的剂量分布及NTCP。
在半扇形模式下,靶区体积的2%和98%所接受的辐射剂量平均分别增加0.90 Gy和0.74 Gy,在全扇形模式下平均分别增加0.76 Gy和0.72 Gy。均匀性指数保持不变。直肠和膀胱在各剂量下受照射的体积百分比略有增加,最大增幅<1%。添加CBCT剂量后,直肠NTCP从0.46%增加到0.53%,增加了约0.07%,而膀胱中的最大NTCP约为0.02%。
本研究展示了一种使用构建的系统评估CBCT和治疗剂量联合剂量的方法。联合剂量分布显示直肠和膀胱剂量的体积增加<1%,直肠NTCP增加约0.07%。