Radiol Phys Technol. 2023 Jun;16(2):203-211. doi: 10.1007/s12194-023-00708-3. Epub 2023 Mar 6.
The use of cone-beam computed tomography (CBCT) is expanding owing to its installation in linear accelerators for radiation therapy, and the imaging dose induced by this system has become the center of attention. Here, the dose to patients caused by the CBCT imager was investigated. Organ doses and effective doses for male and female mesh-type reference computational phantoms (MRCPs) and pelvis CBCT mode, routinely used for pelvic irradiation, were estimated using the Particle and Heavy Ion Transport Code System. The simulation results were confirmed based on the point-dose measurements. The estimated organ doses for male MRCPs with/without raised arms and for female MRCPs with/without raised arms were 0.00286-35.6 mGy, 0.00286-35.1 mGy, 0.00933-39.5 mGy, and 0.00931-39.0 mGy, respectively. The anticipated effective doses for male MRCPs with/without raised arms and female MRCPs with/without raised arms irradiated by pelvis CBCT mode were 4.25 mSv, 4.16 mSv, 7.66 mSv, and 7.48 mSv, respectively. The results of this study will be useful for patients who undergo image-guided radiotherapy with CBCT. However, because this study only covered one type of cancer with one type of imager, and image quality was not considered, more studies should be conducted to estimate the radiation dose from imaging devices in radiation therapy.
锥形束计算机断层扫描(CBCT)的应用因其在放射治疗线性加速器中的安装而不断扩大,该系统引起的成像剂量已成为关注的焦点。本文研究了 CBCT 成像仪对患者的剂量。使用粒子和重离子传输代码系统估算了男性和女性网孔型参考计算体模(MRCP)和骨盆 CBCT 模式的器官剂量和有效剂量,后者通常用于骨盆照射。根据点剂量测量结果对模拟结果进行了验证。对于男性 MRCP 有/无举起手臂和女性 MRCP 有/无举起手臂,估计的器官剂量分别为 0.00286-35.6 mGy、0.00286-35.1 mGy、0.00933-39.5 mGy 和 0.00931-39.0 mGy。骨盆 CBCT 模式照射的男性 MRCP 有/无举起手臂和女性 MRCP 有/无举起手臂的预期有效剂量分别为 4.25 mSv、4.16 mSv、7.66 mSv 和 7.48 mSv。本研究的结果将对接受 CBCT 图像引导放射治疗的患者有用。但是,由于本研究仅涵盖一种癌症和一种成像仪,并且未考虑图像质量,因此应进行更多研究以估算放射治疗中成像设备的辐射剂量。