Suppr超能文献

立体数字减影血管造影术的深度测定

Depth determination on stereoscopic digital subtraction angiograms.

作者信息

Higashida Y, Hirata Y, Saito R, Doudanuki S, Bussaka H, Takahashi M

机构信息

Department of Radiological Technology, College of Medical Science, Kumamoto University, Kuhonji, Japan.

出版信息

Radiology. 1988 Aug;168(2):560-2. doi: 10.1148/radiology.168.2.3293114.

Abstract

The determination of depth on stereoscopic digital subtraction angiographic (DSA) images was evaluated. A "plus" phantom that contained simulated blood vessels of various sizes with various concentrations of contrast medium was used for the study. The vessels ranged in diameter from 0.46 to 1.35 mm, with concentrations of contrast material of 1.6%-25%. The images were evaluated by seven radiologists and six radiologic technologists. The detectability of depth separation increased as the iodine concentration and the vessel size increased. With stereoscopic DSA, correct identification of depth is possible more than 80% of the time for vessels approximately 1 mm in diameter and containing 3.1% contrast medium when the vessels are separated by more than 2.8 mm.

摘要

对立体数字减影血管造影(DSA)图像上深度的测定进行了评估。本研究使用了一个“加号”体模,其包含不同大小、含有不同浓度造影剂的模拟血管。血管直径范围为0.46至1.35毫米,造影剂浓度为1.6% - 25%。七名放射科医生和六名放射技师对这些图像进行了评估。随着碘浓度和血管大小的增加,深度分离的可检测性提高。对于直径约1毫米、含有3.1%造影剂且血管间距超过2.8毫米的情况,使用立体DSA时,超过80%的时间能够正确识别深度。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验