Clin Radiol. 2021 Jan;76(1):77.e17-77.e21. doi: 10.1016/j.crad.2020.08.023. Epub 2020 Sep 17.
To evaluate the radiation dose and diagnostic image quality of low-dose computed tomography (CT) of the paranasal sinus in children, with acquisition at an ultra-low tube voltage (70 kVp) combined with the Flash technique.
Eighty paediatric patients underwent CT of the paranasal sinus and were divided into two groups according to different protocols (group A: 80 kVp protocol with conventional spiral mode [n=40] and group B: 70 kVp protocol with Flash scan mode [n=40]). For each examination, the CT dose index (CTDIvol), dose-length product (DLP), and effective dose (ED) were estimated. The image noise, signal-to-noise ratio (SNR), and overall subjective diagnostic image quality were also evaluated.
For radiation dose, the CTDIvol (mGy), DLP (mGy·cm), and ED (mSv) values of the 70 kVp protocol were significantly lower than those of the 80 kVp protocol (CTDIvol: 1.57±0.009 versus 0.39±0.004 mGy, p<0.001; DLP: 19.88±2.01 versus 6.31±0.52 mGy·cm, p<0.001; ED: 0.079±0.016 versus 0.024±0.005 mSv, p<0.001). Compared with those of the 80-kVp protocol, the image noise increased by 40.7% (p=0.113), the SNR decreased by 48.9%, and the SNR increased by 10.1% with the 70-kVp protocol (p=0.176 and 0.227, respectively). There was no significant difference in the overall subjective image quality grades between these two groups (p=0.15).
When imaging the paranasal sinus in children, an ultra-low tube voltage (70 kVp) combined with the Flash CT technique can reduce the radiation dose significantly while maintaining diagnostic image quality with clinically acceptable image noise.
评估超低管电压(70kVp)联合 Flash 技术在儿童鼻窦低剂量 CT 中的辐射剂量和诊断图像质量。
80 例儿科患者行鼻窦 CT 检查,根据不同的方案分为两组(A 组:80kVp 常规螺旋模式[40 例]和 B 组:70kVp Flash 扫描模式[40 例])。对每例检查,计算 CT 剂量指数(CTDIvol)、剂量长度乘积(DLP)和有效剂量(ED)。还评估了图像噪声、信噪比(SNR)和整体主观诊断图像质量。
在辐射剂量方面,70kVp 方案的 CTDIvol(mGy)、DLP(mGy·cm)和 ED(mSv)值明显低于 80kVp 方案(CTDIvol:1.57±0.009 与 0.39±0.004 mGy,p<0.001;DLP:19.88±2.01 与 6.31±0.52 mGy·cm,p<0.001;ED:0.079±0.016 与 0.024±0.005 mSv,p<0.001)。与 80kVp 方案相比,70kVp 方案的图像噪声增加了 40.7%(p=0.113),SNR 降低了 48.9%,而 SNR 增加了 10.1%(p=0.176 和 0.227)。两组整体主观图像质量评分无显著差异(p=0.15)。
在儿童鼻窦成像中,使用超低管电压(70kVp)联合 Flash CT 技术可以显著降低辐射剂量,同时保持具有临床可接受噪声的诊断图像质量。