HDR brachytherapy well chamber calibration and stability evaluated over twenty years of clinical use.

机构信息

Alfred Health Radiation Oncology, Melbourne, Australia; Australian Radiation Protection and Nuclear Safety Agency, Yallambie, Australia.

Alfred Health Radiation Oncology, Melbourne, Australia.

出版信息

Brachytherapy. 2021 Jan-Feb;20(1):257-264. doi: 10.1016/j.brachy.2020.08.017. Epub 2020 Sep 17.

Abstract

PURPOSE

The purpose of the study was to establish, using a retrospective analysis of existing hospital records, the long-term stability and accuracy of a high-dose-rate brachytherapy well chamber. This should be assessed to determine reliability and appropriate calibration frequency. The accrual of long-term data that demonstrates the stability of our chamber may inform others of the performance they might expect from similar equipment.

METHODS AND MATERIALS

We evaluated air kerma strength measurements made with the PTW 32002 (Nucletron 077.091) high-dose-rate well chamber on 72 Ir sources over an 18-year period and the seven calibrations of that chamber which span a 27-year period.

RESULTS

Consecutive air kerma strength measurements agreed within 0.01% on average. The chamber measurement agreed with the source specification within 0.02% on average, but was up to 1.4% during some calibration periods. The chamber calibration coefficient varied by a maximum of 5% over seven chamber calibration measurements.

CONCLUSIONS

The constancy of the well chamber current compared with the source manufacturer suggests that our chamber has been stable to better than 2% over a period of 18 years. Although the chamber has received different calibration coefficients over time, these coefficients are within the combined uncertainties of any two calibrations and are consistent with the chamber being stable. The agreement we have observed between clinical measurements and the source manufacturer would justify an action level for further investigation of 1%, for this specific chamber.

摘要

目的

本研究旨在通过对现有医院记录的回顾性分析,确定高剂量率近距离治疗后装贮源水箱的长期稳定性和准确性。这应评估以确定可靠性和适当的校准频率。积累长期数据可以证明我们的腔室的稳定性,从而让其他人了解他们可能从类似设备中获得的性能。

方法和材料

我们评估了在 18 年期间使用 PTW 32002(Nucletron 077.091)高剂量率后装贮源水箱对 72Ir 源进行的空气比释动能强度测量,以及跨越 27 年的该水箱的 7 次校准。

结果

连续的空气比释动能强度测量平均相差在 0.01%以内。腔室测量与源规格平均相差在 0.02%以内,但在某些校准期间高达 1.4%。腔室校准系数在 7 次腔室校准测量中最大变化 5%。

结论

与源制造商相比,后装贮源水箱电流的恒定性表明,在 18 年的时间内,我们的腔室稳定性优于 2%。尽管腔室随时间接收不同的校准系数,但这些系数在任何两次校准的综合不确定度内,并且与腔室的稳定性一致。我们观察到临床测量与源制造商之间的一致性,为进一步调查 1%的特定腔室提供了行动水平的依据。