IEEE Trans Med Imaging. 2021 Jan;40(1):205-217. doi: 10.1109/TMI.2020.3024722. Epub 2020 Dec 29.
We present the design and performance of a new compact preclinical system combining positron emission tomography (PET) and magnetic resonance imaging (MRI) for simultaneous scans. The PET contains sixteen SiPM-based detector heads arranged in two octagons and covers an axial field of view (FOV) of 102.5 mm. Depth of interaction effects and detector's temperature variations are compensated by the system. The PET is integrated in a dry magnet operating at 7 T. PET and MRI characteristics were assessed complying with international standards and interferences between both subsystems during simultaneous scans were addressed. For the rat size phantom, the peak noise equivalent count rates (NECR) were 96.4 kcps at 30.2 MBq and 132.3 kcps at 28.4 MBq respectively with and without RF coil. For mouse, the peak NECR was 300.0 kcps at 34.5 MBq and 426.9 kcps at 34.3 MBq respectively with and without coil. At the axial centre of the FOV, spatial resolutions expressed as full width at half maximum / full width at tenth maximum (FWHM/FWTM) ranged from 1.69/3.19 mm to 2.39/4.87 mm. The peak absolute sensitivity obtained with a 250-750 keV energy window was 7.5% with coil and 7.9% without coil. Spill over ratios of the NEMA NU4-2008 image quality (NEMA-IQ) phantom ranged from 0.25 to 0.96 and the percentage of non-uniformity was 5.7%. The image count versus activity was linear up to 40 MBq. The principal magnetic field variation was 0.03 ppm/mm over 40 mm. The qualitative and quantitative aspects of data were preserved during simultaneous scans.
我们提出了一种新的紧凑型临床前系统的设计和性能,该系统结合了正电子发射断层扫描(PET)和磁共振成像(MRI),用于同时扫描。PET 包含十六个基于 SiPM 的探测器头,排列在两个八角形中,覆盖 102.5mm 的轴向视野(FOV)。系统补偿了相互作用深度效应和探测器的温度变化。PET 集成在一个工作在 7T 的干式磁体中。根据国际标准评估了 PET 和 MRI 的特性,并解决了同时扫描时两个子系统之间的干扰问题。对于大鼠大小的体模,在 30.2MBq 和 28.4MBq 时,分别有和没有 RF 线圈,峰值噪声等效计数率(NECR)为 96.4k cps 和 132.3k cps。对于小鼠,在 34.5MBq 和 34.3MBq 时,分别有和没有线圈,峰值 NECR 为 300.0k cps 和 426.9k cps。在 FOV 的轴向中心,空间分辨率表示为半最大值全宽/十分之九全宽(FWHM/FWTM),范围从 1.69/3.19mm 到 2.39/4.87mm。用 250-750keV 能量窗口获得的最大绝对灵敏度为 7.5%,有线圈,7.9%,无线圈。NEMA NU4-2008 图像质量(NEMA-IQ)体模的溢出比范围为 0.25 至 0.96,不均匀性百分比为 5.7%。图像计数与活性成正比,最高可达 40MBq。主磁场变化在 40mm 内为 0.03ppm/mm。同时扫描时,数据的定性和定量方面都得到了保留。