SimPET:用于同时进行 PET/MR 成像的临床前 PET 插入设备。

SimPET: a Preclinical PET Insert for Simultaneous PET/MR Imaging.

机构信息

Brightonix Imaging Inc., Yeonmujang 5ga-gil, Seongdong-gu, Seoul, 04782, South Korea.

Department of Biomedical Sciences, Seoul National University College of Medicine, 103 Daehak-ro, Jongno-gu, Seoul, 03080, South Korea.

出版信息

Mol Imaging Biol. 2020 Oct;22(5):1208-1217. doi: 10.1007/s11307-020-01491-y.

Abstract

PURPOSE

SimPET/M7 system is a small-animal dedicated simultaneous positron emission tomography and magnetic resonance imaging (PET/MRI) scanner. The SimPET insert has been upgraded from its prototype with a focus on count rate performance and sensitivity. The M7 scanner is a 1-T permanent magnet-based compact MRI system without any cryogens. Here, we present performance evaluation results of SimPET along with the results of mutual interference evaluation and simultaneously acquired PET/MR imaging.

PROCEDURES

Following NEMA NU 4-2008 standard, we evaluated the performance of the SimPET system. The M7 MRI compatibility of SimPET was also assessed by analyzing MRI images of a uniform phantom under different PET conditions and PET count rates with different MRI pulse sequences. Mouse imaging was performed including a whole-body F-NaF PET scan and a simultaneous PET/MRI scan with Cu-NOTA-ironoxide.

RESULTS

The spatial resolution at center based on 3D OSEM without and with warm background was 0.7 mm and 1.45 mm, respectively. Peak sensitivity was 4.21 % (energy window = 250-750 keV). The peak noise equivalent count rate with the same energy window was 151 kcps at 38.4 MBq. The uniformity was 4.42 %, and the spillover ratios in water- and air-filled chambers were 14.6 % and 12.7 %, respectively. In the hot rod phantom image, 0.75-mm-diameter rods were distinguishable. There were no remarkable differences in the SNR and uniformity of MRI images and PET count rates with different PET conditions and MRI pulse sequences. In the whole-body F-NaF PET images, fine skeletal structures were well resolved. In the simultaneous PET/MRI study with Cu-NOTA-ironoxide, both PET and MRI signals changed before and after injection of the dual-modal imaging probe, which was evident with the exact spatiotemporal correlation.

CONCLUSIONS

We demonstrated that the SimPET scanner has a high count rate performance and excellent spatial resolution. The combined SimPET/M7 enabled simultaneous PET/MR imaging studies with no remarkable mutual interference between the two imaging modalities.

摘要

目的

SimPET/M7 系统是一款小型动物专用正电子发射断层扫描和磁共振成像(PET/MRI)扫描仪。SimPET 插件已经从原型进行了升级,重点是提高计数率性能和灵敏度。M7 扫描仪是一种基于 1 特斯拉永磁体的紧凑型 MRI 系统,不使用任何制冷剂。在此,我们展示了 SimPET 的性能评估结果,以及相互干扰评估和同时进行的 PET/MR 成像结果。

程序

根据 NEMA NU 4-2008 标准,我们评估了 SimPET 系统的性能。还通过分析不同 PET 条件下的均匀体模的 MRI 图像和不同 MRI 脉冲序列下的 PET 计数率,评估了 SimPET 对 M7 MRI 的兼容性。进行了小鼠成像,包括全身 F-NaF PET 扫描和带有 Cu-NOTA-氧化铁的同时 PET/MRI 扫描。

结果

基于 3D OSEM 的中心空间分辨率在没有和有暖背景的情况下分别为 0.7 毫米和 1.45 毫米。峰值灵敏度为 4.21%(能量窗=250-750keV)。相同能量窗的峰值噪声等效计数率为 151kcps,在 38.4MBq 时。均匀性为 4.42%,水填充和空气填充腔的溢出比分别为 14.6%和 12.7%。在热棒体模图像中,可分辨出 0.75 毫米直径的棒。在不同的 PET 条件和 MRI 脉冲序列下,MRI 图像的 SNR 和均匀性以及 PET 计数率没有明显差异。在全身 F-NaF PET 图像中,精细的骨骼结构得到很好的分辨。在带有 Cu-NOTA-氧化铁的同时 PET/MRI 研究中,在注射双模态成像探针前后,PET 和 MRI 信号都发生了变化,这与精确的时空相关性是一致的。

结论

我们证明了 SimPET 扫描仪具有高计数率性能和优异的空间分辨率。组合的 SimPET/M7 能够进行同时的 PET/MR 成像研究,两种成像模式之间没有明显的相互干扰。