Comparison of phacoemulsification parameters in eyes with and without exfoliation syndrome.

作者信息

Tulu Aygun B, Altan C, Kirmaci Kabakci A

机构信息

University of Health Sciences, Beyoglu Eye Training and Research Hospital, Bereketzade Street Number:2, Istanbul, Turkey.

University of Health Sciences, Beyoglu Eye Training and Research Hospital, Bereketzade Street Number:2, Istanbul, Turkey.

出版信息

J Fr Ophtalmol. 2020 Dec;43(10):1031-1038. doi: 10.1016/j.jfo.2020.02.010. Epub 2020 Sep 21.

Abstract

OBJECTIVE

To evaluate and to compare the phacoemulsification machine parameters in eyes with and without XFS, requiring cataract surgery.

METHODS

Patients who underwent phacoemulsification and in-the-bag IOL implantation for cataract were included in this retrospective study. All surgeries were performed by the same experienced surgeon using the stop & chop technique with the same phacoemulsification device (Infiniti Vision System, Alcon Laboratories, Inc., USA). Patients were divided into two groups according to the presence of exfoliation material (XFM). Each group consisted of consecutive patients. Their characteristics and intraoperative phacoemulsification parameters were compared.

RESULTS

Sixty-eight eyes of 68 patients [29 in the exfoliation syndrome (XFS) (-) group, 39 in XFS (+)] were enrolled. There were no statistical differences regarding preoperative patient characteristics. There was a statistically significant difference in total U/S time, phaco time, aspiration time and estimated fluid used between the XFS (+) and XFS (-) groups (P=0.021, P=0.017, P=0.009 and P=0.002, respectively). Considering that the use of a CTR (capsule tension ring) might be an important factor potentially affecting surgical parameters, the data were analyzed accordingly. Aspiration time and estimated fluid used remained statistically significant (P=0.046 and P=0.017, respectively); however, although the U/S total and phaco time were found to be longer in XFS (+) group compared to XFS (-) group, the difference did not show statistical significance (P=0.061 and P=0.059, respectively). There were no differences between groups regarding endothelial cell loss or any other postoperative complications.

CONCLUSIONS

The presence of XFS results in longer total U/S time, phaco and aspiration time and more estimated fluid used in phacoemulsification, but this prolongation does not result in additional complications.

摘要

目的

评估并比较需要进行白内障手术的伴有和不伴有假性剥脱综合征(XFS)的眼中超声乳化仪参数。

方法

本回顾性研究纳入了接受白内障超声乳化及囊袋内人工晶状体植入术的患者。所有手术均由同一位经验丰富的外科医生使用相同的超声乳化设备(美国爱尔康公司的Infiniti Vision System)采用拦截劈核技术完成。根据是否存在剥脱物质(XFM)将患者分为两组。每组由连续的患者组成。比较两组患者的特征及术中超声乳化参数。

结果

纳入了68例患者的68只眼[剥脱综合征(XFS)阴性组29只眼,XFS阳性组39只眼]。术前患者特征方面无统计学差异。XFS阳性组与XFS阴性组在总超声时间、超声乳化时间、抽吸时间及估计液体用量方面存在统计学显著差异(分别为P = 0.021、P = 0.017、P = 0.009和P = 0.002)。考虑到使用囊袋张力环(CTR)可能是潜在影响手术参数的重要因素,因此对数据进行了相应分析。抽吸时间和估计液体用量仍具有统计学显著性(分别为P = 0.046和P = 0.017);然而,尽管发现XFS阳性组的总超声时间和超声乳化时间比XFS阴性组更长,但差异无统计学显著性(分别为P = 0.061和P = 0.059)。两组在内皮细胞丢失或任何其他术后并发症方面无差异。

结论

XFS的存在导致超声乳化术中总超声时间、超声乳化和抽吸时间更长,且估计液体用量更多,但这种延长并未导致额外的并发症。