Departments of1Neurosurgery and.
2Orthopaedics and Physical Medicine, Medical University of South Carolina, Charleston, South Carolina.
J Neurosurg Pediatr. 2020 Oct 2;26(6):667-670. doi: 10.3171/2020.6.PEDS20299. Print 2020 Dec 1.
Although the advent of magnetic growing rod technology for scoliosis has provided a means to bypass multiple hardware lengthening operations, it is important to be aware that many of these same patients have a codiagnosis of hydrocephalus with magnet-sensitive programmable ventricular shunts. As the magnetic distraction of scoliosis rods has not previously been described to affect the shunt valve setting, the authors conducted an investigation to characterize the interaction between the two devices.
In this ex vivo study, the authors carried out 360 encounters between four different shunt valve types at varying distances from the magnetic external remote control (ERC) used to distract the growing rods. Valve settings were examined before and after every interaction with the remote control to determine if there was a change in the setting.
The Medtronic Strata and Codman Hakim valves were found to have setting changes at distances of 3 and 6 inches but not at 12 inches. The Aesculap proGAV and Codman Certas valves, typically described as MRI-resistant, did not have any setting changes due to the magnetic ERC regardless of distance.
Although it is not necessary to check a shunt valve after every magnetic distraction of scoliosis growing rods, if there is concern that the magnetic ERC may have been within 12 inches (30 cm) of a programmable ventricular shunt valve, the valve should be checked at the bedside with a programmer or with a skull radiograph along with postdistraction scoliosis radiographs.
尽管矫形脊柱侧凸的磁生长棒技术的出现为避免多次硬件延长操作提供了一种手段,但重要的是要意识到,许多此类患者还伴有脑积水和磁敏感程控脑室分流器。由于脊柱侧凸杆的磁牵引以前并未被描述为会影响分流器阀门的设置,因此作者进行了一项研究以描述这两种设备之间的相互作用。
在这项离体研究中,作者在距离用于牵引生长棒的外部磁遥控器(ERC)不同距离处对四种不同的分流器阀门进行了 360 次接触。在与遥控器进行每次交互后检查阀门设置,以确定设置是否发生变化。
发现 Medtronic Strata 和 Codman Hakim 阀门在距离 3 英寸和 6 英寸处有设置变化,但在 12 英寸处没有变化。通常被描述为 MRI 耐受的 Aesculap proGAV 和 Codman Certas 阀门无论距离如何,由于磁 ERC,均未发生任何设置变化。
尽管在每次对脊柱侧凸生长棒进行磁牵引后不必检查分流器阀门,但如果担心磁 ERC 可能已在程控脑室分流器阀门的 12 英寸(30 厘米)范围内,则应在床边使用程控器或颅骨射线照相以及后牵引脊柱侧凸射线照相进行检查。