Hossain Md-Billal, Lazaro Jesus, Noh Yeonsik, Chon Ki H
Annu Int Conf IEEE Eng Med Biol Soc. 2020 Jul;2020:592-595. doi: 10.1109/EMBC44109.2020.9175665.
We propose a novel electrocardiogram (ECG) denoising technique using the variable frequency complex demodulation (VFCDM) algorithm. We used VFCDM to perform the sub-band decomposition of the noise-contaminated ECG to remove the noise components so that accurate QRS complexes could be identified. The ECG quality was further improved by removing baseline drift and smoothing via adaptive mean filtering. The proposed method was validated on the MIT-BIH arrhythmia database (MITDB) and wearable armband ECG data. For the former, we added Gaussian white noise to the ECG signals at different signal-to-noise ratios and the denoising performance of the proposed method was compared with other denoising techniques. The proposed approach showed superior denoising performance compared to the other methods. We compared the QRS complex detection performance of the noisy to the denoised armband ECG. The performance of the proposed denoising method on the armband ECG resulted in comparable QRS complex detection as that obtained when using Holter monitor ECG signals. This demonstrates that the proposed algorithm can significantly increase the amount of usable armband ECG data, which would otherwise have been discarded due to electromyogram contamination especially during arm movements. Hence, the proposed algorithm has the potential to enable long-term monitoring of atrial fibrillation using the armband without the discomfort of skin irritation often experienced with Holter monitors.Clinical Relevance- The proposed ECG denoising method can significantly increase the ECG quality of wearable ECG devices, which are more susceptible to muscle and motion artifacts.
我们提出了一种使用可变频率复解调(VFCDM)算法的新型心电图(ECG)去噪技术。我们使用VFCDM对受噪声污染的ECG进行子带分解,以去除噪声成分,从而能够识别准确的QRS复合波。通过去除基线漂移并采用自适应均值滤波进行平滑处理,进一步提高了ECG质量。所提出的方法在麻省理工学院-贝斯以色列女执事医疗中心心律失常数据库(MITDB)和可穿戴臂带ECG数据上得到了验证。对于前者,我们以不同的信噪比向ECG信号中添加高斯白噪声,并将所提出方法的去噪性能与其他去噪技术进行了比较。与其他方法相比,所提出的方法显示出卓越的去噪性能。我们比较了有噪声的臂带ECG与去噪后的臂带ECG的QRS复合波检测性能。所提出的去噪方法在臂带ECG上的性能导致QRS复合波检测结果与使用动态心电图监测仪ECG信号时相当。这表明所提出的算法可以显著增加可用的臂带ECG数据量,否则这些数据会因肌电图干扰而被丢弃,尤其是在手臂运动期间。因此,所提出的算法有潜力使用臂带对房颤进行长期监测,而不会出现动态心电图监测仪常有的皮肤刺激不适。临床相关性——所提出的ECG去噪方法可以显著提高可穿戴ECG设备的ECG质量,这类设备更容易受到肌肉和运动伪影的影响。