Performance assessment of two motion management systems for frameless stereotactic radiosurgery.

机构信息

Department of Radiation Oncology, Shanghai Chest Hospital, Shanghai Jiao Tong University, Shanghai, China.

Department of Radiation Oncology, Rutgers-Cancer Institute of New Jersey, Rutgers-Robert Wood Johnson Medical School, 195 Little Albany St., New Brunswick, NJ, USA.

出版信息

Strahlenther Onkol. 2021 Feb;197(2):150-157. doi: 10.1007/s00066-020-01688-8. Epub 2020 Oct 12.

Abstract

BACKGROUND/PURPOSE: Frameless stereotactic radiosurgery (SRS) requires dedicated systems to monitor patient motion in order to avoid inaccurate radiation delivery due to involuntary shifts. The purpose of this study is to assess the accuracy and sensitivity of two distinct motion monitoring systems used for frameless SRS.

METHODS

A surface image-guided system known as optical surface monitoring system (OSMS), and a fiducial marker-based system known as high definition motion management (HDMM) as part of the latest Gamma Knife Icon® were compared. A 3D printer-based cranial motion phantom was developed to evaluate the accuracy and sensitivity of these two systems in terms of: (1) the capability to recognize predefined shifts up to 3 cm, and (2) the capability to recognize predefined speeds up to 3 cm/s. The performance of OSMS, in terms of different reference surfaces, was also evaluated.

RESULTS

Translational motion could be accurately detected by both systems, with an accuracy of 0.3 mm for displacement up to 1 cm, and 0.5 mm for larger displacements. The reference surface selection had an impact on OSMS performance, with flat surface resulting in less accuracy. HDMM was in general more sensitive when compared with OSMS in capturing the motion, due to its faster frame rate, but a delay in response was observed with faster speeds. Both systems were less sensitive in detection of superior-inferior motion when compared to lateral or vertical displacement directions.

CONCLUSION

Translational motion can be accurately and sensitively detected by OSMS and HDMM real-time monitoring systems. However, performance variations were observed along different motion directions, as well as amongst the selection of reference images. Caution is needed when using real-time monitoring systems for frameless SRS treatment.

摘要

背景/目的:无框架立体定向放射外科(SRS)需要专用系统来监测患者的运动,以避免由于无意识的移动而导致辐射输送不准确。本研究的目的是评估两种用于无框架 SRS 的不同运动监测系统的准确性和灵敏度。

方法

比较了一种称为光学表面监测系统(OSMS)的表面图像引导系统和一种基于基准标记的系统,称为高清运动管理(HDMM),作为最新的 Gamma Knife Icon®的一部分。开发了一个基于 3D 打印机的颅运动体模,以评估这两种系统在以下方面的准确性和灵敏度:(1)识别高达 3cm 的预定移动的能力,以及(2)识别高达 3cm/s 的预定速度的能力。还评估了 OSMS 在不同参考表面方面的性能。

结果

两种系统都能准确地检测到平移运动,对于 1cm 以内的位移,精度为 0.3mm,对于更大的位移,精度为 0.5mm。参考表面的选择对 OSMS 的性能有影响,平面表面的精度较低。与 OSMS 相比,HDMM 在捕获运动方面通常更敏感,因为它的帧率更快,但在更快的速度下观察到响应延迟。与侧向或垂直位移方向相比,这两种系统在检测上下运动时的灵敏度较低。

结论

OSMS 和 HDMM 实时监测系统可以准确和敏感地检测平移运动。然而,在不同的运动方向以及参考图像的选择上观察到了性能的变化。在进行无框架 SRS 治疗时,需要谨慎使用实时监测系统。