Suppr超能文献

Ti(IV)修饰的乙烯膦酸磁性纳米粒子用于从鸡肉样品中同时预富集多种砷形态,然后用 HPLC-ICP-MS 进行分析。

Ti (IV) modified vinyl phosphate magnetic nanoparticles for simultaneous preconcentration of multiple arsenic species from chicken samples followed by HPLC-ICP-MS analysis.

机构信息

Key Laboratory of Analytical Chemistry for Biology and Medicine (Ministry of Education), Department of Chemistry, Wuhan University, Wuhan, Hubei Province, 430072, P. R. China.

出版信息

Electrophoresis. 2021 Feb;42(4):465-472. doi: 10.1002/elps.202000158.

Abstract

Ti (IV)-modified vinyl phosphate magnetic nanoparticles (Fe O @SiO @KH570-PO -Ti (IV)) was prepared for simultaneous extraction of multiple arsenic species, followed by high performance liquid chromatography (HPLC)- inductively coupled plasma mass spectrometry (ICP-MS) analysis. Inorganic arsenic (iAs), dimethyl arsenic acid (DMA), monomethyl arsenic acid (MMA), p-amino phenyl arsenic acid (p-ASA), 4-hdroxyphenylarsenic acid (4-OH), phenyl arsenic acid (PAA), and 3-nitro-4-hydroxyphenylarsenic acid (ROX) were investigated as interest analytes. It was found that they were quantitatively adsorbed on Fe O @SiO @KH570-PO -Ti (IV) at pH 5, and desorbed completely with 0.1 mol/L sodium hydroxide solution. Enrichment factor of 100-fold was obtained by consuming 100 mL sample solution. Under the optimal conditions, the method combining MSPE with HPLC-ICP-MS presented a linear range of 1-5000 ng/L for seven arsenic species. The limits of detection were 0.39, 0.60, 0.23, 1.85, 0.54, 0.48, and 0.84 ng/L for DMA, MMA, p-ASA, iAs, 4-OH, PAA, ROX, with the relative standard deviations (c = 10 ng/L, n = 7) of 3.6, 3.9, 5.5, 12.4, 6.1, 5.8, 5.0, respectively. The accuracy of the method was validated by analyzing BCR 627 Tuna fish. The application potential of the method was further evaluated by chicken muscle and liver samples. No target arsenic species were detected in these samples, and good recoveries (80.6-123%) were obtained for the spiked samples at low, medium, and high concentration levels.

摘要

Ti(IV)- 改性乙烯基膦酸磁性纳米粒子(Fe3O4@SiO2@KH570-PO4-Ti(IV))被制备用于同时提取多种砷形态,随后通过高效液相色谱(HPLC)-电感耦合等离子体质谱(ICP-MS)分析。无机砷(iAs)、二甲基砷酸(DMA)、一甲基砷酸(MMA)、对氨基苯砷酸(p-ASA)、4- 羟基苯砷酸(4-OH)、苯砷酸(PAA)和 3- 硝基-4- 羟基苯砷酸(ROX)被作为感兴趣的分析物进行研究。结果发现,它们在 pH 5 时被定量吸附在 Fe3O4@SiO2@KH570-PO4-Ti(IV)上,并可完全用 0.1 mol/L 氢氧化钠溶液洗脱。通过消耗 100 mL 样品溶液,可获得 100 倍的富集因子。在最佳条件下,MSPE 与 HPLC-ICP-MS 相结合的方法对七种砷形态的线性范围为 1-5000 ng/L。DMA、MMA、p-ASA、iAs、4-OH、PAA、ROX 的检出限分别为 0.39、0.60、0.23、1.85、0.54、0.48 和 0.84 ng/L,相对标准偏差(c = 10 ng/L,n = 7)分别为 3.6、3.9、5.5、12.4、6.1、5.8、5.0。通过分析 BCR 627 金枪鱼对该方法的准确性进行了验证。通过鸡肉和肝脏样品进一步评估了该方法的应用潜力。在这些样品中未检测到目标砷形态,并且在低、中、高浓度水平下,对添加的样品均获得了良好的回收率(80.6-123%)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验