Abstract

SIGNIFICANCE

Measuring pupil diameter may provide an additional objective method to detect and monitor glaucoma.

PURPOSE

The purposes of this study were to evaluate whether pupillometry can be used in the assessment of glaucomatous damage and to determine whether pupillometer can be used in glaucoma screening.

METHODS

In this retrospective study, patients with early stages of glaucomatous optic neuropathy were tested using visual evoked potential (VEP), standard automated perimetry, and pupillometer. The VEP record was made using two consecutive check sizes (1° and 15') at 1-Hz frequency. Patients with a mean deviation of <-2 and >-6 dB were included in the study. Pupil size measurements were performed under photopic conditions at 60-, 100-, and 130-cd/m luminance levels with the OPD Scan (Nidek, Gamagori, Japan) device.

RESULTS

In all, 24 eyes of 24 patients (14 women and 10 men) with newly diagnosed open-angle glaucoma were included, and 30 eyes of 30 healthy subjects (15 women, 15 men) were used as controls. There was no significant difference between sexes in photopic pupil diameters in either group. Comparing the pupil diameters in different luminance levels, the diameters were significantly larger in the patient group. When age-corrected values in the glaucoma group were compared with an age-matched control group, the differences in P100 latency and amplitudes were statistically significant. There were significant correlations between photopic pupil diameters and timing of VEP waveforms.

CONCLUSIONS

There is a relationship between pupil diameters and VEP results in eyes with early glaucoma.

摘要

意义

测量瞳孔直径可能提供一种额外的客观方法来检测和监测青光眼。

目的

本研究旨在评估瞳孔计是否可用于评估青光眼损害,并确定瞳孔计是否可用于青光眼筛查。

方法

在这项回顾性研究中,使用视觉诱发电位(VEP)、标准自动视野计和瞳孔计对早期青光眼视神经病变患者进行了测试。VEP 记录使用两个连续的检查尺寸(1°和 15')在 1-Hz 频率下进行。包括平均偏差为<-2 和>-6 dB 的患者。在 60-、100-和 130-cd/m 亮度水平下,使用 OPD Scan(日本 Nidek,Gamagori)设备进行瞳孔大小测量。

结果

共纳入 24 例新诊断的开角型青光眼患者的 24 只眼(14 名女性和 10 名男性),30 例健康受试者的 30 只眼(15 名女性,15 名男性)作为对照组。两组中,在明适应瞳孔直径方面,性别之间无显著差异。比较不同亮度水平下的瞳孔直径,患者组的直径明显较大。当将青光眼组的年龄校正值与年龄匹配的对照组进行比较时,P100 潜伏期和振幅的差异具有统计学意义。明适应瞳孔直径与 VEP 波形的时间之间存在显著相关性。

结论

在早期青光眼的眼中,瞳孔直径与 VEP 结果之间存在关系。