Sarezky Daniel, Volpe Nicholas J, Park Meghan S, Tanna Angelo P
Department of Ophthalmology, Northwestern University Feinberg School of Medicine, Chicago, IL.
J Glaucoma. 2016 Mar;25(3):312-6. doi: 10.1097/IJG.0000000000000213.
To assess the correlations between the relative afferent pupillary defect measured by a binocular, computerized, and commercially available pupillometer and the inter-eye differences in average retinal nerve fiber layer (RNFL) thickness measured by spectral-domain optical coherence tomography (SD-OCT) and visual field index (VFI) and mean deviation (MD) measured by automated static perimetry in patients with glaucoma.
Cross-sectional study.
Thirty four patients with open-angle glaucoma.
Pupillary constriction amplitudes were measured with an automated pupillometer. SD-OCT imaging of the circumpapillary RNFL and automated static perimetry were performed within 6 months of pupillometry.
Main outcome measures included Spearman correlation coefficients between relative afferent pupillary defect as measured by pupillometry and inter-eye differences in average RNFL thickness, VFI, and MD.
Correlation coefficients between relative afferent pupillary defect and inter-eye difference in average RNFL thickness, MD, and VFI were -0.81 (P<0.001), -0.73 (P<0.001), and -0.68 (P<0.001), respectively.
The automated pupillometer quantifies asymmetry in afferent pupillary responses in proportion to structural and functional asymmetry as measured SD-OCT and automated static perimetry.
评估使用双目、计算机化且市售的瞳孔计测量的相对传入性瞳孔障碍与通过光谱域光学相干断层扫描(SD-OCT)测量的双眼平均视网膜神经纤维层(RNFL)厚度差异、视野指数(VFI)以及通过自动静态视野计测量的平均偏差(MD)之间的相关性,这些研究对象为青光眼患者。
横断面研究。
34例开角型青光眼患者。
使用自动瞳孔计测量瞳孔收缩幅度。在瞳孔测量后6个月内进行视乳头周围RNFL的SD-OCT成像和自动静态视野检查。
主要观察指标包括瞳孔计测量的相对传入性瞳孔障碍与双眼平均RNFL厚度差异、VFI和MD之间的Spearman相关系数。
相对传入性瞳孔障碍与双眼平均RNFL厚度差异、MD和VFI之间的相关系数分别为-0.81(P<0.001)、-0.73(P<0.001)和-0.68(P<0.001)。
自动瞳孔计根据SD-OCT和自动静态视野计测量的结构和功能不对称性,按比例量化传入性瞳孔反应的不对称性。