Novel oxygen saturation imaging endoscopy to assess anastomotic integrity in a porcine ischemia model.

机构信息

Department of Colorectal Surgery, National Cancer Center Hospital East, 6-5-1 Kashiwanoha, Kashiwa, Chiba, 277-8577, Japan.

Surgical Device Innovation Office, NEXT Medical Device Innovation Center, National Cancer Center Hospital East, Kashiwa, Japan.

出版信息

BMC Surg. 2020 Oct 22;20(1):250. doi: 10.1186/s12893-020-00913-6.

Abstract

BACKGROUND

Establishing anastomotic integrity is crucial for avoiding anastomotic complications in colorectal surgery. This study aimed to evaluate the safety and feasibility of assessing anastomotic integrity using novel oxygen saturation imaging endoscopy in a porcine ischemia model.

METHODS

In three pigs, a new endoscope system was used to check the mechanical completeness of the anastomosis and capture the tissue oxygen saturation (StO) images. This technology can derive the StO images from the differences in the absorption coefficient in the visible light region between oxy- and deoxy-hemoglobin. Bowel perfusion at the proximal rectum was assessed before and after the anastomosis, and 1 min and 30 min after the ligation of the cranial rectal artery (CRA).

RESULTS

The completeness of the anastomoses was confirmed by the absence of air leakage. Intraluminal oxygen saturation imaging was successfully performed in all animals. There was no significant difference in the StO level before and after the anastomosis (52.6 ± 2.0 vs. 52.0 ± 2.6; p = 0.76, respectively). The StO level of the intestine on the oral side of the anastomosis one minute after the CRA ligation was significantly lower than immediately after the anastomosis (15.9 ± 6.0 vs. 52.0 ± 2.6; p = 0.006, respectively). There was no significant difference in the StO level between 1 min after and 30 min after the CRA ligation (15.9 ± 6.0 vs. 12.1 ± 5.3; p = 0.41, respectively).

CONCLUSION

Novel oxygen saturation imaging endoscopy was safe and feasible to assess the anastomotic integrity in the experimental model.

摘要

背景

在结直肠手术中,建立吻合完整性对于避免吻合口并发症至关重要。本研究旨在评估新型氧饱和度成像内镜在猪缺血模型中评估吻合完整性的安全性和可行性。

方法

在三只猪中,使用新型内镜系统检查吻合口的机械完整性并获取组织氧饱和度(StO)图像。该技术可以从可见光谱区域中氧合血红蛋白和去氧血红蛋白之间的吸收系数差异中得出 StO 图像。在吻合前和吻合后、直肠近端结扎颅状直肠动脉(CRA)后 1 分钟和 30 分钟评估肠灌注情况。

结果

所有动物的吻合完整性均通过无空气泄漏来确认。所有动物均成功进行了腔内氧饱和度成像。吻合前后 StO 水平无显著差异(52.6±2.0 与 52.0±2.6;p=0.76)。CRA 结扎后 1 分钟时吻合口口腔侧肠段的 StO 水平明显低于吻合后即刻(15.9±6.0 与 52.0±2.6;p=0.006)。CRA 结扎后 1 分钟与 30 分钟时的 StO 水平无显著差异(15.9±6.0 与 12.1±5.3;p=0.41)。

结论

新型氧饱和度成像内镜安全且可行,可用于评估实验模型中的吻合完整性。